FDA Adverse Event Malfunction Summary report: N

PLATE HOLDING DRILL GUIDE - SINGLE BARREL/FIXED ANGLE

MDR report key: 2925555 · Received January 22, 2013

Report

Report Number
8030965-2013-00221
Event Type
Malfunction
Date Received
January 22, 2013
Date of Event
December 31, 2012
Report Date
December 31, 2012
Manufacturer
SYNTHES GMBH
Product Code
FZX
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THIS DEVICE IS USED FOR TREATMENT, NOT DIAGNOSIS. SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS. THE MANUFACTURING DOCUMENTS WERE REVIEWED AND NO COMPLAINT RELATED ISSUES WERE FOUND.

Additional Manufacturer Narrative · 1

DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED/EXPLANTED. A PRODUCT DEVELOPMENT EVALUATION WAS CONDUCTED AND THE REPORT INDICATES THE DRILL GUIDE WAS RECEIVED WITH A BROKEN HANDLE AND OXIDATION ON THE ETCHING. THERE ARE ALSO MINOR SCRATCHES ON THE SURFACE. THE HANDLE IS FRACTURED ABOVE THE PIN WHICH RETAINS THE RADEL HANDLE ON THE MAIN BODY OF THE INSTRUMENT. THE FAILURE COULD HAVE OCCURRED FROM EITHER THE INSTRUMENT BEING DROPPED ON THE FLOOR OR COMPRESSED BEYOND THE INTENDED FUNCTION FOR PLATE RETAINMENT. AS THE FRACTURED HANDLE OCCURRED OUTSIDE OF THE SURGICAL FIELD AND NOT AT THE BACK TABLE IT IS NOT CLEAR HOW IT BROKE. THE ENGINEER LOCATED THE RISK ANALYSIS FOR THIS PART AND THE RISK OF THE HANDLE BREAKING IS NOT ADDRESSED.

Description of Event or Problem · 1

PRIOR TO AN ACDF AT C5-C6, THE VECTRA DRILL GUIDE BROKE. THIS HAPPENED IN THE OPERATING ROOM, BUT OUTSIDE THE SURGICAL FIELD, NOT AT THE BACK TABLE. THERE WERE NO FRAGMENTS TO RETRIEVE AND THERE WAS NO HARM TO THE PATIENT. ANOTHER DRILL GUIDE WAS AVAILABLE TO USE. THE SURGEON COMPLETED THE PROCEDURE WITH NO FURTHER PROBLEM AND THE PATIENT IS REPORTEDLY RECOVERING WELL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
29442 PLATE HOLDING DRILL GUIDE - SINGLE BARREL/FIXED ANGLE PLATE HOLDING DRILL GUIDE FZX SYNTHES GMBH 1597473

Patients

Seq Age Sex Outcome Treatment
1 47 YR