TRIMA ACCEL
Report
- Report Number
- 1722028-2013-00025
- Event Type
- Malfunction
- Date Received
- January 22, 2013
- Date of Event
- December 24, 2012
- Report Date
- December 26, 2012
- Manufacturer
- TERUMO BCT
- Product Code
- GKT
- PMA / PMN Number
- BK120017
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
INVESTIGATION: THE RUN DATA FILE (RDF) WAS ANALYZED FOR THIS EVENT. THE SIGNALS IN THE RUN DATAFILE INDICATE THAT THE HIGHER THAN EXPECTED WBC CONTENT IN THE PLATELET PRODUCT WAS LIKELY A RESULT OF AN ESCAPE OF WBCS FROM THE LRS CHAMBER DURING A PORTION OF THE PROCEDURE. THERE ARE NO EVENTS (ADJUSTMENTS, CHANGES IN PUMP SPEED, SUBSTATE CHANGES, ETC.) IN THE PROCEDURE THAT CORRESPONDS WITH THE ONSET OF THE OVERLOADING OF THE LRS CHAMBER. BASED ON THE AVAILABLE INFORMATION, IT CANNOT BE RULED OUT THAT THIS LEUKOREDUCTION FAILURE COULD ALSO BE DONOR-RELATED. INVESTIGATION EVALUATION AND CORRECTIVE ACTIONS ARE IN-PROCESS. A FOLLOW-UP REPORT WILL BE PROVIDED.
INVESTIGATION: THE DEVICE HISTORY RECORDS (DHR) WERE REVIEWED FOR THIS LOT. THERE WERE NO EVENTS NOTED IN THE DHR THAT WOULD HAVE CONTRIBUTED TO THE ELEVATED WBC COUNT EXPERIENCED BY THE CUSTOMER. ROOT CAUSE: THIS DISPOSABLE SET WAS UNAVAILABLE FOR SPECIFIC ROOT CAUSE ANALYSIS. A DEFINITIVE ROOT CAUSE FOR THE OBSERVED LEUKOREDUCTION FAILURE REMAINS UNDETERMINED AT THIS TIME. POSSIBLE ROOT CAUSES WERE PROVIDED IN THE INITIAL REPORT FOR THIS EVENT.
THE CUSTOMER WOULD LIKE THE RUN DATA FILE INVESTIGATED TO DETERMINE A POSSIBLE CAUSE FOR THE ELEVATED WHITE BLOOD CELL (WBC) CONTENT IN THE PLATELET PRODUCT. THERE WAS NOT A TRANSFUSION RECIPIENT OR PATIENT INVOLVED AT THE TIME OF THE RESIDUAL WBCL TESTING, THEREFORE, NO PATIENT INFORMATION IS REASONABLY KNOWN AT THE TIME OF THE EVENT. DONOR UNIT# (B)(4). THE DISPOSABLE KIT IS NOT AVAILABLE FOR RETURN BECAUSE IT WAS DISCARDED BY THE CUSTOMER. THIS REPORT IS BEING FILED DUE TO A DEVICE MALFUNCTION THAT HAS THE POTENTIAL FOR INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 30709 | TRIMA ACCEL | TRIMA ACCEL PLATELET, PLASMA SET | GKT | TERUMO BCT | 11U2111 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |