FDA Adverse Event
Malfunction
Summary report: N
840 VENTILATOR
MDR report key: 2925355
·
Received January 17, 2013
Report
- Report Number
- 8020893-2013-00126
- Event Type
- Malfunction
- Date Received
- January 17, 2013
- Date of Event
- January 2, 2013
- Report Date
- January 2, 2013
- Manufacturer
- COVIDIEN, FORMERLY NELLCOR PURITAN
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
THE SERVICE REPORT SHOWS THE CUSTOMER REPORTED THAT THE 840 VENTILATOR UPPER DISPLAY IS ERRATIC. THERE WAS NO PT INVOLVEMENT AT THE TIME OF THE EVENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE GRAPHICAL USER INTERFACE (GUI) PCB. THE UNIT PASSED EXTENDED SELF TESTING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 26844 | 840 VENTILATOR | VENTILATOR | CBK | COVIDIEN, FORMERLY NELLCOR PURITAN | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |