FDA Adverse Event Injury Summary report: N

PFC*LINER 28MM/ 54-58MM,10 DEG

MDR report key: 2925317 · Received January 22, 2013

Report

Report Number
1818910-2013-11310
Event Type
Injury
Date Received
January 22, 2013
Date of Event
January 9, 2013
Report Date
January 9, 2013
Manufacturer
DEPUY ORTHOPAEDICS, INC.
Product Code
JDI
PMA / PMN Number
PK873585
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
ME, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

PATIENT WAS REVISED TO ADDRESS OSTEOLYSIS AND POLY WEAR. LOOSENING OF THE STEM AT THE CEMENT/BONE INTERFACE WAS ALSO REPORTED; HOWEVER, COMPETITOR CEMENT WAS USED AT THE TIME OF ORIGINAL IMPLANTATION. DOI 1997 - DOR (B)(6) 2013 (HIP). THE DEVICE ASSOCIATED WITH THIS REPORT WAS NOT RETURNED. REVIEW OF THE DEVICE HISTORY RECORDS AND/OR A LOT SPECIFIC COMPLAINT DATABASE SEARCH WAS NOT POSSIBLE AS THE LOT CODE REQUIRED WAS NOT PROVIDED. REQUESTS FOR ADDITIONAL INVESTIGATIONAL INPUTS WERE MADE IN ACCORDANCE WITH WI-7915 APPENDIX A; REV. C. NO ADDITIONAL INFORMATION WAS OBTAINED. THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS REGARDING THE REPORTED EVENT WITHOUT DEVICE EXAMINATION. HOWEVER ALTHOUGH NOT RETURNED, IT WOULD NOT BE UNREASONABLE TO FIND POLY MATERIAL WEAR AFTER THE LENGTH OF TIME REPORTED AS IMPLANTED. BASED ON THE INVESTIGATION THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT OR ADDITIONAL INFORMATION BE RECEIVED TO CHANGE THE OUTCOME OF THE PERFORMED INVESTIGATION, THE COMPLAINT WILL BE RE-OPENED.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Description of Event or Problem · 1

PATIENT WAS REVISED TO ADDRESS OSTEOLYSIS AND POLY WEAR. LOOSENING OF THE STEM AT THE CEMENT/BONE INTERFACE WAS ALSO REPORTED; HOWEVER, COMPETITOR CEMENT WAS USED AT THE TIME OF ORIGINAL IMPLANTATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
28896 PFC*LINER 28MM/ 54-58MM,10 DEG ACETAUBLAR LINER JDI DEPUY ORTHOPAEDICS, INC. UNKNOWN

Patients

Seq Age Sex Outcome Treatment
1 83 YR Required Intervention