FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 2925290 · Received January 17, 2013

Report

Report Number
8020893-2013-00119
Event Type
Malfunction
Date Received
January 17, 2013
Date of Event
December 18, 2012
Report Date
December 18, 2012
Manufacturer
COVIDIEN, FORMERLY NELLCOR PURITAN BENNETT
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

RECEIVED INFORMATION STATING THAT DUE TO A VENTILATOR MALFUNCTION PT WAS PLACED ON A SECOND VENTILATOR. PT WAS NOT HARMED OR INJURED AS A RESULT OF THE EVENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (TSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE OXYGEN SENSOR. THE UNIT PASSED EXTENDED SELF TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
26658 840 VENTILATOR VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR PURITAN BENNETT 840

Patients

Seq Age Sex Outcome Treatment
1