FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2925277 · Received January 22, 2013

Report

Report Number
2531779-2013-01005
Event Type
Malfunction
Date Received
January 22, 2013
Date of Event
December 28, 2012
Report Date
December 28, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS: INVESTIGATION REVEALED THAT THE FORCE SENSOR WAS OUT OF CALIBRATION AND THE FORCE RESISTANCE MEASUREMENT IS OUT OF SPECIFICATION. A REVIEW OF THE TOTAL DAILY DOSE HISTORY INDICATED THAT INSULIN DELIVERY TOTALS CORRECTLY REFLECTED PROGRAMMED VALUES. AN EZPRIME OPERATION WAS PERFORMED WITH NO ISSUES OCCURRING. THE PUMP WAS EXERCISED FOR (B)(4) WITH NO DELIVERY ISSUES. THE PUMP WAS EVALUATED AND FOUND TO BE OPERATING WITHIN REQUIRED SPECIFICATIONS AND DELIVERING INSULIN ACCURATELY. UNRELATED TO THE FORCE SENSOR ISSUE, EVALUATION REVEALED THE INTERNAL CLOCK BATTERY ON THE PCB BOARD HAD FAILED. THE PUMP WOULD NOT RETAIN THE USER PROGRAMMED DATE AND TIME SETTINGS UPON REMOVAL OF THE PRIMARY AA BATTERY. WHEN A NEW AA BATTERY IS INSERTED THE PUMP DISPLAYS THE DEFAULT DATE AND TIME WHICH MUST BE MANUALLY CONFIRMED (OR RESET) BY THE USER IN ORDER TO PROCEED. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT THE FORCE SENSOR WAS OUT OF CALIBRATION AND THE FORCE RESISTANCE MEASUREMENT IS OUT OF SPECIFICATION. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2012.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
31124 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 19 YR