OT VERIO IQ METER
Report
- Report Number
- 3008382007-2013-01284
- Event Type
- Malfunction
- Date Received
- January 22, 2013
- Report Date
- December 27, 2012
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K110637
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4): THE REPORTED COMPLAINT WAS OBSERVED ON THE ERROR LOG, HOWEVER, NO MESSAGE WAS OBSERVED DURING CONTROL SOLUTION TESTING. PA WAS UNABLE TO REPRODUCE THE COMPLAINT. THE PATIENT ALSO RETURNED AN EMPTY VIAL OF TEST STRIPS; PA IS UNABLE TO DO ANY EVALUATION. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
THE PATIENT'S PRODUCTS HAVE BEEN RETURNED TO LIFESCAN FOR EVALUATION; HOWEVER, THE EVALUATION PROCESS HAS NOT YET BEEN COMPLETED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2012, THE LAY USER/PATIENT CONTACTED LIFESCAN TO REPORT THE ONE TOUCH VERIO IQ METER WAS GIVING THE ERROR 2 ERROR MESSAGE. THERE WAS NO PATIENT INJURY ASSOCIATED WITH THIS ISSUE. AS THE ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING, THIS COMPLAINT IS BEING REPORTED. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 30424 | OT VERIO IQ METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3305559 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |