OT VERIO IQ METER
Report
- Report Number
- 3008382007-2013-01283
- Event Type
- Malfunction
- Date Received
- January 22, 2013
- Report Date
- December 27, 2012
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K110637
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP # 1 (03/05/2013)-DEVICE EVALUATION: THE METER INVOLVED WITH THIS COMPLAINT HAS BEEN RETURNED ON (B)(4) 2013 AND EVALUATED BY PRODUCT ANALYSIS (PA) ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE METER PASSED TESTING WITH NO FAULTS FOUND. NO ERROR MESSAGE WAS OBSERVED. THE METER FUNCTIONED PROPERLY. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.
THE LAY USER/PATIENT'S PRODUCT(S) HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE TEST STRIPS INVOLVED IN THIS CASE WERE TESTED AND FOUND TO HAVE FAILED GIVING ERROR 4 DURING TESTING. IF ANY ADDITIONAL INFORMATION IS AVAILABLE, THE FDA WILL BE NOTIFIED IN A FOLLOW UP REPORT. AT THIS TIME, WE CONSIDER THIS MATTER CLOSED.
ON (B)(6) 2012 THE LAY USER/PATIENT CONTACTED LIFESCAN TO REPORT THE ONE TOUCH VERIO IQ METER WAS GIVING THE ERROR 4 ERROR MESSAGE. THERE WAS NO PATIENT INJURY ASSOCIATED WITH THIS ISSUE. AS THE ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING, THIS COMPLAINT IS BEING REPORTED. THE LAY USER/PATIENT'S PRODUCT(S) HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS WITH THE FOLLOWING FINDINGS: THE TEST STRIPS INVOLVED IN THIS CASE WERE TESTED AND FOUND TO HAVE FAILED GIVING ERROR 4 DURING TESTING. IF ANY ADDITIONAL INFORMATION IS AVAILABLE, THE FDA WILL BE NOTIFIED IN A FOLLOW UP REPORT. AT THIS TIME, WE CONSIDER THIS MATTER CLOSED.