HOMECHOICE PRO
Report
- Report Number
- 1416980-2013-01609
- Event Type
- Malfunction
- Date Received
- January 22, 2013
- Date of Event
- December 4, 2012
- Report Date
- January 7, 2013
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- SERVICE AND TESTING PERSONNEL
Narratives
(B)(4). (B)(6). THE DEVICE WAS RETURNED TO BAXTER, AND THE EVALUATION IS COMPLETE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE REPORTED DIFFICULTY OF AN IIPV EVENT WAS CONFIRMED THROUGH AN EVENT HISTORY LOG REVIEW. THE CAUSE WAS UNDETERMINED. IF ANY RELEVANT ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.
DURING THE EVALUATION OF A RETURNED HOMECHOICE MACHINE, AN INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED DURING THE THERAPY INITIATED ON (B)(6) 2012 AT 08:57:54. DURING NIGHT DRAIN CYCLE FIVE, THE PATIENT'S ULTRAFILTRATION READING WAS 887ML, INDICATING THE HOME PATIENT (HP) DRAINED 887ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 600ML. THIS INFORMATION MEETS IIPV CRITERIA. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 30591 | HOMECHOICE PRO | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |