FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2924860 · Received January 22, 2013

Report

Report Number
2531779-2013-00974
Event Type
Malfunction
Date Received
January 22, 2013
Report Date
December 26, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 (B)(4). DEVICE HISTORY RECORD REVIEW WAS CONDUCTED ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS. THE BUTTON SYMBOLS WERE FOUND TO BE WORN BUT THE KEYPAD WAS FOUND TO BE INTACT. THE UP, DOWN, AND CONTRAST BUTTONS WERE FOUND TO BE INTERMITTENTLY RESPONSIVE. THE KEYPAD WAS REMOVED AND CONTAMINATION WAS OBSERVED UNDER ALL OF THE BUTTON CONTACTS. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED TO ANIMAS. INVESTIGATION REVEALED CONTAMINATION UNDER THE BUTTON CONTACTS. THIS REPORT IS MADE BASED ON THE RESULTS OF INVESTIGATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
30476 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 11 YR