ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 2924860
·
Received January 22, 2013
Report
- Report Number
- 2531779-2013-00974
- Event Type
- Malfunction
- Date Received
- January 22, 2013
- Report Date
- December 26, 2012
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
FOLLOW-UP # 1 (B)(4). DEVICE HISTORY RECORD REVIEW WAS CONDUCTED ON (B)(4) 2012 WITH THE FOLLOWING FINDINGS. THE BUTTON SYMBOLS WERE FOUND TO BE WORN BUT THE KEYPAD WAS FOUND TO BE INTACT. THE UP, DOWN, AND CONTRAST BUTTONS WERE FOUND TO BE INTERMITTENTLY RESPONSIVE. THE KEYPAD WAS REMOVED AND CONTAMINATION WAS OBSERVED UNDER ALL OF THE BUTTON CONTACTS. (B)(6).
Description of Event or Problem · 1
THE PUMP WAS RETURNED TO ANIMAS. INVESTIGATION REVEALED CONTAMINATION UNDER THE BUTTON CONTACTS. THIS REPORT IS MADE BASED ON THE RESULTS OF INVESTIGATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 30476 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |