FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 2924714 · Received January 22, 2013

Report

Report Number
3004209178-2013-90469
Event Type
Malfunction
Date Received
January 22, 2013
Date of Event
December 17, 2012
Report Date
January 11, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP ALARMED DURING THE BASIC OCCLUSION TEST, AND THE DEVICE WAS UNABLE TO PRIME DURING THE PRIME TEST AS A RESULT OF A LOOSE DRIVE SUPPORT DISK.

Description of Event or Problem · 1

THE CUSTOMER REPORTED REGARDING ISSUES DURING PRIME/REWIND. THE BLOOD GLUCOSE READING WAS 376MG/DL. THE CALLER STATED THAT THE INSULIN PUMP ALARMED DURING THE MANUAL PRIME PROCESS. THE CUSTOMER MENTIONED ALSO THAT THE DEVICE ALARMED MOTOR ERROR. THE CALLER STATED THAT THE INSULIN PUMP WAS NOT EXPOSED TO A HIGH MAGNETIC FIELD. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
30261 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAB

Patients

Seq Age Sex Outcome Treatment
1 42 YR