FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2924689 · Received January 22, 2013

Report

Report Number
3004209178-2013-90463
Event Type
Malfunction
Date Received
January 22, 2013
Date of Event
December 4, 2012
Report Date
January 11, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP ALARMED DURING THE PRIME TEST, AND THE DEVICE ALARMED MOTOR ERROR DURING THE BASIC OCCLUSION TEST AS A RESULT OF A LOOSE DRIVE SUPPORT DISK.

Description of Event or Problem · 1

THE CUSTOMER REPORTED REGARDING ISSUES DURING PRIME/REWIND. THE BLOOD GLUCOSE READING WAS 160MG/DL. THE CALLER STATED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. THE INFUSION SET AND RESERVOIR WERE CHANGED. THE CALLER STATED WHILE PRIMING THE INSULIN SQUIRTED OUT DURING, AND THE DEVICE ALARMED. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NO FURTHER INFORMATION WAS PROVIDED.