UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER
Report
- Report Number
- 2050012-2013-00049
- Event Type
- Malfunction
- Date Received
- January 22, 2013
- Date of Event
- December 28, 2012
- Report Date
- December 28, 2012
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
A BEC FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED FOR THIS EVENT. THE FSE IDENTIFIED THAT ALL THE LEAKING FLUID WAS CONTAINED BY THE DESIGNED DRIP CHANNELS AND GRAVITY DRAIN. THE FSE INDICATED THAT THE CARTRIDGE CHEMISTRY (CC) WASH COLLAR WAS NOT FULLY INSERTED INTO THE WASH COLLAR HOLDER AND WAS TOO LOW; MEANING THAT NO VACUUM WAS SEALED TO THE WASH COLLAR. THE FSE REPLACED THE VACUUM VALVE, THE CC SAMPLE PROBE, AND THE WASH COLLAR, WHICH RESOLVED THE ISSUE. SYSTEM PERFORMANCE WAS ALSO VERIFIED.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) REPORTING A CONTAINED CARTRIDGE CHEMISTRY (CC) SAMPLE PROBE LEAK IN THE UNICEL DXC 800 PRO SYNCHRON CHEMISTRY ANALYZER. THE LEAK WAS OBSERVED FROM THE CC SAMPLE PROBE WASH COLLAR WHILE RUNNING SAMPLES. AS THE CC SAMPLE PROBE WAS BEING WASHED, FLUID WOULD SQUIRT FROM THE WASH COLLAR. THE CUSTOMER INDICATED THAT APPROXIMATELY 15 MLS LEAKED FROM THE INSTRUMENT. THE CUSTOMER WAS WEARING PROPER PROTECTIVE EQUIPMENT (PPE), INCLUDING GLOVES AND LABORATORY COAT WHEN THE LEAK WAS DISCOVERED. THERE WAS NO CONTACT TO OPEN OR BROKEN SKIN OR MUCOUS MEMBRANES. NO ONE SOUGHT MEDICAL ATTENTION. THE CUSTOMER INDICATED THAT FIVE (5) PATIENTS GENERATED ERRONEOUS RESULTS ON MULTIPLE TESTS DUE TO THE LEAK. THE PATIENT SAMPLES WERE RERUN ON AN ALTERNATE BACKUP INSTRUMENT IN THE CUSTOMER'S LABORATORY. THE ERRONEOUS RESULTS WERE REPORTED OUT OF THE LABORATORY; HOWEVER, AMENDED REPORTS WERE SUBMITTED. THE RESULTS PROVIDED BY THE CUSTOMER ARE SHOWN IN THIS REPORT. THERE WAS NO REPORT OF IMPACT TO PATIENT TREATMENT AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 30903 | UNICEL® DXC 800 PRO SYNCHRON® CHEMISTRY ANALYZER | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JJE | BECKMAN COULTER INC. | DXC 800 PRO | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |