DRILL BIT Ø1 L50/35

MDR report key: 2924492 · Received January 22, 2013

Report

Report Number
1719045-2013-00160
Event Type
Malfunction
Date Received
January 22, 2013
Date of Event
December 6, 2012
Report Date
December 26, 2012
Manufacturer
SYNTHES MONUMENT
Product Code
HTW
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED/EXPLANTED. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS THE PRODUCT IS ENTERING THE COMPLAINT SYSTEM. A REVIEW OF THE DEVICE HISTORY RECORDS HAS BEEN REQUESTED.

Additional Manufacturer Narrative · 1

ORCHID UNIQUE MANUFACTURED THE 1.0MM DRILL BIT MINI QC 50MM, PART 316.396, LOT U145568. THE LOTS WERE REVIEWED AND NO NON CONFORMITIES WERE FOUND. THE LOT PASSED SYNTHES INSPECTION SHEET. THE BROKEN SURFACE IS HOMOGENOUS, WHAT INDICATES MATERIAL CONFORMITY AS WELL. THE DRILL BIT IS BROKEN DUE TO MECHANICAL OVERLOADING DURING USE. WE STRONGLY HAVE TO ASSUME THAT THE TIP OF THE DRILL BITS CAME IN STRONG CONTACT WITH THE K WIRE AND BLOCKED WHAT LEAD TO THE BREAKAGE FINALLY. NO PRODUCT FAULT COULD BE DETECTED.

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES EUROPE REPORTS AN EVENT IN (B)(6) AS FOLLOWS: DURING A RADIAL HEAD FRACTURE PROCEDURE USING HCS AND MHS PLATE, TWO DRILL BITS WERE BROKEN. THE SURGEON REMOVED EVERY VISIBLE BROKEN PIECE FROM THE SURGICAL FIELD WITH DIFFICULTY AND THE PROCEDURE WAS COMPLETED USING ANOTHER DRILL BIT. THE SURGEON REPORTEDLY DID FEEL THAT HE PUT EXTRA LOAD ON THE GUIDE WIRE. IT WAS REPORTED THE PATIENT HAD A GOOD BONE QUALITY. THIS IS 1 OF 2 REPORTS FOR THE SAME EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
29990 DRILL BIT Ø1 L50/35 DRILL BIT HTW SYNTHES MONUMENT U145568

Patients

Seq Age Sex Outcome Treatment
1 28 YR DRILL BIT