FDA Adverse Event Malfunction Summary report: N

DHS/DCS COUPLING SCREW

MDR report key: 2924482 · Received January 22, 2013

Report

Report Number
2530088-2013-00055
Event Type
Malfunction
Date Received
January 22, 2013
Date of Event
December 26, 2012
Report Date
December 26, 2012
Manufacturer
SYNTHES BRANDYWINE
Product Code
JDO
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

.

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED/EXPLANTED. PRODUCT DEVELOPMENT ENGINEER EVALUATED THE DEVICES AND INDICATED THAT THE COMPLAINT CONDITION WAS MOST LIKELY CAUSED BY AN OFF-AXIS FORCE BEING APPLIED TO THE RETURNED DHS/DCS COUPLING SCREW (PART#338.20) WHICH CAUSED THE THREADED PORTION TO SHEAR OFF. THE COUPLING SCREW WAS MANUFACTURED IN AUGUST 2008 AND IS OVER 4 YEARS OLD. THE DEVICE IS MANUFACTURED FROM 17-4PH MATERIAL WHICH IS A TYPICAL MATERIAL SYNTHES USES TO MAKE CANNULATED COUPLING SCREWS. THE RETURNED DHS/DCS GUIDE SHAFT (PART#338.21) WAS MANUFACTURED IN SEPTEMBER 1998 AND IS OVER 14 YEARS OLD. BASED ON THE AGE OF THE RETURNED DEVICE, THIS DEVICE HAS OUTLIVED ITS USEFUL LIFE. THE MANUFACTURE DATE OF RETURNED DHS/DCS LAG SCREW (PART#280.950) WAS MANUFACTURED IN SEPTEMBER 2012. THE DEVICE IS MANUFACTURED FROM 316L SS WHICH IS A TYPICAL MATERIAL SYNTHES USES TO MAKE LAG SCREWS. THE COMPLAINT CONDITION WAS CAUSED BY AN OFF AXIS FORCE CAUSING THE THREAD TO FRACTURE AT THE LOCATION JUST ABOVE WHERE THE THREAD WAS SUPPORTED INSIDE THE RETURNED LAG SCREW. THE RISK ANALYSIS WAS REVIEWED AND IT DID NOT ADDRESS THE COMPLAINT CONDITION FOR THE THREADS SHEARING DUE TO OFF-AXIS FORCE. PRODUCT DEVELOPMENT WAS NOTIFIED TO UPDATE/RECREATE RISK ANALYSIS TO ADEQUATELY ADDRESS THESE COMPLAINT CONDITIONS. A REVIEW OF THE DEVICE HISTORY RECORD REVEALED NO COMPLAINT RELATED ANOMALIES.

Description of Event or Problem · 1

DURING AN INTERTROCHANTERIC PROXIMAL FEMUR FRACTURE PROCEDURE USING DHS SYSTEM, SOME OF THE THREADS OF THE INSERTER BROKE OFF INTO THE LAG SCREW DURING INSERTION. ALL FRAGMENTS WERE REPORTEDLY RETRIEVED. THE SURGEON REMOVED THE LAG SCREW AND INSERTER, USED A DIFFERENT LAG SCREW AND INSERTER, AND COMPLETED THE PROCEDURE. THERE WAS NO EXTENSION OF SURGERY TIME. THIS IS 1 OF 3 REPORTS FOR THE SAME EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
30381 DHS/DCS COUPLING SCREW DHS/DCS COUPLING SCREW JDO SYNTHES BRANDYWINE 5746101

Patients

Seq Age Sex Outcome Treatment
1 81 YR GUIDE SHAFT, LAG SCREW