FDA Adverse Event Malfunction Summary report: N

OT VERIO IQ METER

MDR report key: 2924360 · Received January 22, 2013

Report

Report Number
3008382007-2013-01186
Event Type
Malfunction
Date Received
January 22, 2013
Report Date
December 28, 2012
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K110637
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

((B)(4) 2013)-DEVICE EVALUATION: THE METER, TEST STRIPS AND CONTROL SOLUTION INVOLVED WITH THIS COMPLAINT HAVE BEEN RETURNED ON (B)(4) 2013 AND EVALUATED BY PRODUCT ANALYSIS (PA) ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE METER PASSED ALL TESTING WITH NO FAULTS FOUND. THE CONTROL SOLUTION DID NOT REQUIRE TESTING. THE TEST STRIPS HAVE NOT YET COMPLETED TESTING, THEREFORE, NO CONCLUSIONS CAN BE DRAWN AT THIS TIME. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A SECOND FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Additional Manufacturer Narrative · 1

FOLLOW-UP (3/2/2013)-DEVICE EVALUATION: THE METER AND TEST STRIPS INVOLVED WITH THIS COMPLAINT HAVE BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON (B)(4) 2013 AND (B)(4) 2013 RESPECTIVELY WITH THE FOLLOWING FINDINGS: THE METER PASSED ALL TESTING WITH NO FAULTS FOUND. THE ERROR COULD NOT BE REPRODUCED. THE TEST STRIPS WERE ALSO TESTED AND THE PRIMARY COMPLAINT WAS NOT CONFIRMED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED. DEVICE RETURNED TO MFG DATE: TEST STRIPS - (B)(6) 2013, METER - (B)(6) 2013.

Description of Event or Problem · 1

ON (B)(6) 2012, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING BLOOD READING AS CONTROL. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
29916 OT VERIO IQ METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3326379

Patients

Seq Age Sex Outcome Treatment
1