FDA Adverse Event Malfunction Summary report: N

AUTOPULSE NIMH BATTERY

MDR report key: 2924182 · Received December 14, 2012

Report

Report Number
3003793491-2012-00376
Event Type
Malfunction
Date Received
December 14, 2012
Date of Event
April 17, 2012
Report Date
April 17, 2012
Manufacturer
ZOLL CIRCULATION, INC.
Product Code
DRM
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

BATTERY SN (B)(4) WAS NOT RETURNED, BUT TEST DATA WAS PROVIDED. BATTERY MAINTENANCE PROGRAM LITERATURE INDICATES THAT THE USEFUL LIFE OF EACH AUTOPULSE BATTERY IS BETWEEN 2-4 YEARS. WITHIN THE 2-4 YEAR TIME PERIOD, THE LIFE OF EACH BATTERY IS INFLUENCED BY THE FREQUENCY OF USE AN THE FREQUENCY OF ROUTINE BATTERY MAINTENANCE. BASED ON THE MANUFACTURED DATE (12/2009), THIS BATTERY MAY HAVE AGED PAST THE END OF ITS USEFUL LIFE. HOWEVER, BATTERY DATA SHOWED A FULL CHARGE AFTER TEST CYCLE. THEREFORE, THE EXPERIENCED EVENT MAY BE ATTRIBUTED TO NOT PERFORMING DAILY SWAPS AND/OR NOT FULLY CHARGING THE BATTERY BEFORE DEPLOYMENT NO ADVERSE EVENT REPORTED.

Description of Event or Problem · 1

FOR (B)(4), IT WAS REPORTED THAT 49 BATTERIES WERE NOT OPERATING CORRECTLY. FOR BATTERY SN (B)(4) REPORTED DESCRIPTION WAS THAT THE BATTERY DID NOT WORK EVEN THOUGH IT WAS FULLY CHARGED. NO ADVERSE EVENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AUTOPULSE NIMH BATTERY NIMH BATTERY DRM ZOLL CIRCULATION, INC.

Patients

Seq Age Sex Outcome Treatment
1 Other