FDA Adverse Event Malfunction Summary report: N

PROGRASP (TM) FORCEPS

MDR report key: 2924100 · Received January 22, 2013

Report

Report Number
2955842-2013-00254
Event Type
Malfunction
Date Received
January 22, 2013
Date of Event
December 20, 2012
Report Date
December 27, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT ONE GRIP CLOSE CABLE WAS FOUND TO BE DERAILED AT THE IDLER PULLEY. THE GRIPS WERE STILL ABLE TO OPEN AND CLOSE WITH CLICKING SNAP SOUNDS, BUT THEIR MOVEMENT COULD NOT BE PRECISE. THE CABLE DERAILMENT WAS LIKELY DUE TO THE CABLE LOSING CONTACT WITH THE PULLEY DURING WRIST ARTICULATION. FLEET ANGLE BETWEEN PROXIMAL AND DISTAL PULLEYS COULD HAD CONTRIBUTED TO THE CABLE DERAILMENT. ADDITIONAL OBSERVATION NOT REPORTED BY THE SITE WAS MAIN TUBE INSULATION DAMAGE. THE DISTAL END OF THE MAIN INSULATION TUBE EXHIBITED VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL, SUGGESTING THAT THEY MAY HAVE BEEN CAUSED BY INSTRUMENT COLLISIONS OR ROUGH HANDLING. THE ENDOWRIST INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DAVINCI SI HYSTERECTOMY PROCEDURE, THE CABLE SNAPPED ON THE PROGRASP FORCEPS INSTRUMENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
31145 PROGRASP (TM) FORCEPS ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420093-08 M10121107 749

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES