PWRD ECH FLEX 60MM
Report
- Report Number
- 3005075853-2013-00289
- Event Type
- Malfunction
- Date Received
- January 22, 2013
- Date of Event
- November 19, 2012
- Report Date
- December 3, 2012
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K110385
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). DAMAGED RELEASE BUTTON. ADDITIONAL INFORMATION WAS REQUESTED AND THE FOLLOWING WAS OBTAINED: ON WHAT TISSUE TYPE WAS THE DEVICE USED? AT WHAT LOCATION ON THE TISSUE?- STOMACH, LOWER TO MIDDLE PORTION. WHAT COLOR CARTRIDGE WAS BEING USED?- GOLD. WHAT OTHER COLOR CARTRIDGES WERE USED BEFORE AND AFTER THIS EVENT? - GREEN. WAS BUTTRESSING MATERIAL UTILIZED? IF SO, WHICH PRODUCT? -NO. WAS THE INSTRUMENT FIRED ACROSS OR NEAR AN EXISTING STAPLE LINE OR CLIP?- YES. A STAPLE LINE. WERE ANY UNEXPECTED NOISES HEARD? IF SO, WHEN?- NO. WERE ANY OF THE FORCES HIGHER OR LOWER THAN EXPECTED (CLOSING, FIRING, OR OPENING)?- NO. AFTER USE, DID EACH OF THE TRIGGERS AND BUTTONS AUTOMATICALLY RETURN TO THEIR ORIGINAL (PRE-FIRED) POSITIONS, WITHOUT INTERVENTION?- NO. WAS THERE ANY DIFFICULTY REMOVING THE DEVICE FROM THE TISSUE? YES. THE ANALYSIS FOUND THAT ONE DEVICE WAS RECEIVED WITHOUT A RELOAD PRESENT AND WITH THE CLAMPING MECHANISM DAMAGED. THE INSTRUMENT WAS TESTED FOR FUNCTIONALITY IN THE STRAIGHT POSITION WITH A TEST CARTRIDGE RELOAD AND ACHIEVED ITS COMPLETE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE DEVICE CLOSED, CYCLED AND OPENED WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. THE DEVICE WAS DISASSEMBLED TO VERIFY THE INTEGRITY OF THE INTERNAL COMPONENTS. UPON DISASSEMBLING, THE CLAMP ARM WAS NOTED TO BE WORN OUT; IT APPEARS THAT THE DEVICE WAS FORCED OPEN WHEN THE KNIFE WAS NOT ON THE HOME POSITION. THE EVENT COULD NOT BE CONFIRMED AS NO CARTRIDGE RELOAD WAS RETURNED FOR ANALYSIS. PLEASE NOTE IF THE INSTRUMENT IS PARTIALLY FIRED, SLIDE THE KNIFE REVERSE SWITCH FORWARD TO RETURN THE KNIFE TO HOME POSITION. TO OPEN THE JAWS, SQUEEZE THE CLOSING TRIGGER, AND THEN SIMULTANEOUSLY PRESS THE ANVIL RELEASE SWITCH ON EITHER SIDE OF THE INSTRUMENT. WHILE PRESSURE IS STILL ON THE ANVIL RELEASE SWITCH, SLOWLY RELEASE THE CLOSING TRIGGER. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING THE SLEEVE GASTRECTOMY, THE STAPLER COMPLETED TWO FIRINGS PERFECTLY. THE STAPLER LOCKED UP ON THE THIRD FIRING. NEW STAPLER WAS OPENED AND THE CASE WAS COMPLETED WITH NO PATIENT CONSEQUENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 30226 | PWRD ECH FLEX 60MM | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | J4C256 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |