FDA Adverse Event Injury Summary report: N

ASPHERE M SPEC 12/14 36 +1.5

MDR report key: 2923803 · Received January 22, 2013

Report

Report Number
1818910-2013-11223
Event Type
Injury
Date Received
January 22, 2013
Date of Event
January 11, 2013
Report Date
March 7, 2014
Manufacturer
DEPUY ORTHOPAEDICS, INC. 1818910
Product Code
KWA
PMA / PMN Number
PK082585
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. A COMPLAINT DATABASE SEARCH FINDS NO OTHER REPORTED INCIDENTS AGAINST THE PROVIDED PRODUCT AND LOT COMBINATION. REVIEW OF THE DEVICE HISTORY RECORDS AND/OR A LOT SPECIFIC COMPLAINT DATABASE SEARCH WAS NOT POSSIBLE FOR THE METAL LINER AS THE PRODUCT AND LOT CODE REQUIRED WAS NOT PROVIDED. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO IDENTIFY ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Additional Manufacturer Narrative · 1

THIS CORRECTION: MANUFACTURING FACILITY: (B)(4). EXAMINATION OF THE REPORTED DEVICES WAS NOT POSSIBLE AS THEY WERE NOT RETURNED. A SEARCH OF THE COMPLAINTS DATABASES IDENTIFIED OTHER REPORTS AGAINST THE FEMORAL HEAD AND INSERT. PER PROCEDURE, THIS DEVICE(S) IS EXEMPT FROM DEVICE HISTORY RECORD REVIEW. A SEARCH OF THE COMPLAINTS DATABASES IDENTIFIED NO OTHER REPORTS AGAINST THE REMAINING PRODUCT/LOT CODE COMBINATIONS. MEDICAL RECORDS WERE RECEIVED AND REVIEWED. THE INVESTIGATION CAN DRAW NO CONCLUSIONS WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO DETERMINE ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION HAS NOT BEEN INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Additional Manufacturer Narrative · 1

LITIGATION PAPERS RECEIVED. IN ADDITION TO WHAT WAS PREVIOUSLY REPORTED, LITIGATION ALLEGES THE PATIENT SUFFERS FROM DISCOMFORT AND TOXIC COBALT-CHROMIUM METAL IONS AND PARTICLES TO BE RELEASED INTO THE BLOOD, TISSUE AND BONE. WE ARE ADDING THE LINER TO ADDRESS THE METAL ON METAL ALLEGATIONS. DEPUY STILL CONSIDERS THE INVESTIGATION CLOSED IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

PATIENT WAS REVISED TO ADDRESS PAIN. DOI: (B)(6) 2010; DOR: (B)(6) 2013 (RIGHT HIP). THE DEVICE ASSOCIATED WITH THIS REPORT WAS NOT RETURNED. A COMPLAINT DATABASE SEARCH FINDS NO OTHER REPORTED INCIDENTS AGAINST THE PROVIDED PRODUCT AND LOT COMBINATION SINCE ITS RELEASE FOR DISTRIBUTION. FOLLOW UP FOR ADDITIONAL INVESTIGATIONAL INPUTS WERE MADE BUT NO ADDITIONAL INFORMATION WAS OBTAINED. THE INVESTIGATION COULD NOT DRAW ANY CONCLUSIONS REGARDING THE REPORTED EVENT. BASED ON THE INABILITY TO DETERMINE A ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT OR ADDITIONAL INFORMATION BE RECEIVED TO CHANGE THE OUTCOME OF THE PERFORMED INVESTIGATION, THE COMPLAINT WILL BE RE-OPENED.

Additional Manufacturer Narrative · 1

DEPUY STILL CONSIDERS THIS INVESTIGATION CLOSED AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

PATIENT WAS REVISED TO ADDRESS PAIN.

Description of Event or Problem · 1

UPDATE 7/29/15-PFS AND MEDICAL RECORDS RECEIVED. AFTER REVIEW OF THE MEDICAL RECORDS FOR MDR REPORTABILITY, THE REVISION OPERATIVE NOTE INDICATED PAIN, DISCOMFORT, METALLOSIS, METAL DEBRIS, AND INCREASED METAL IONS. THE SURGEON WAS ALSO UNABLE TO REMOVE THE LINER SO THE WHOLE CUP WAS REMOVED. THE SURGEON SPENT A TOTAL OF 65 MINUTES REMOVING THE LINER. THE CUP AND STEM ARE BEING ADDED AND THE PART/LOT IS BEING UPDATED FOR THE LINER. THE COMPLAINT WAS UPDATED ON:8/26/2015.

Description of Event or Problem · 1

PPF ALLEGES PSEUDOTUMOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
29424 ASPHERE M SPEC 12/14 36 +1.5 FEMORAL HEAD KWA DEPUY ORTHOPAEDICS, INC. 1818910 3079968

Patients

Seq Age Sex Outcome Treatment
1 62 YR Required Intervention