FDA Adverse Event Injury Summary report: N

12MM/130 DEG TI CANN TROCH FIXATION NAIL 380MM/RIGHT-STER

MDR report key: 2923636 · Received January 22, 2013

Report

Report Number
1719045-2013-00154
Event Type
Injury
Date Received
January 22, 2013
Date of Event
December 25, 2012
Report Date
December 25, 2012
Manufacturer
SYNTHES MONUMENT
Product Code
HSB
PMA / PMN Number
K011857
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
VA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THIS DEVICE IS USED FOR TREATMENT, NOT DIAGNOSIS. THE LOCKING MECHANISM TAB HAS SHEARED OFF AND THE REMAINING PORTION IS BENT INWARD. DURING THIS EVALUATION, THE RETURNED DEVICES WERE ASSEMBLED WITH KNOWN GOOD MATING INSTRUMENTS AND IMPLANTS TO COMPLETE THE INSERTION CONSTRUCT IN ORDER TO EVALUATE THE ALIGNMENT OF THE RETURNED DEVICES. THE CONSTRUCT ASSEMBLED AND THE RETURNED HELICAL BLADE ALIGNED AND ENTERED THE RETURNED TFN AS INTENDED BUT MET RESISTANCE WITH THE DAMAGED/BENT LOCKING MECHANISM. THE COMPLAINT CONDITION WAS REPLICATED. THE RISK ANALYSIS ADEQUATELY ADDRESSES THIS COMPLAINT CONDITION AS LESS SEVERE WITH A MODERATE SEVERITY OF HARM 3 AND AN IMPROBABLE PROBABILITY OF OCCURRENCE 1. THE DESIGN IS ADEQUATE FOR ITS INTENDED USE AND DID NOT CONTRIBUTE TO THIS COMPLAINT CONDITION. IT CAN NOT BE DEFINITIVELY DETERMINED HOW OR WHEN THE LOCKING MECHANISM WAS DEPLOYED.

Additional Manufacturer Narrative · 1

INVESTIGATION COULD NOT BE COMPLETED, NO CONCLUSION COULD BE DRAWN AS NO DEVICE WAS RETURNED. A REVIEW OF SYNTHES DEVICE HISTORY RECORDS FOR MANUFACTURING REVEALED NO COMPLAINT RELATED ISSUES.

Description of Event or Problem · 1

DURING SURGERY FOR A TFN NAILING PROCEDURE ON (B)(6) 2012, THE HELICAL BLADE WOULD NOT ADVANCE THROUGH THE NAIL. THE SURGEON DETERMINED THE LOCKING MECHANISM HAD BEEN DEPLOYED. THE SURGEON REMOVED THE BLADE AND TRIED TO BACK OUT THE LOCKING MECHANISM IN THE NAIL BUT WAS UNABLE TO DO SO BECAUSE THE LOCKING TAB HAD BEEN DAMAGED AND BENT WHILE TRYING TO INSERT THE BLADE. THE NAIL WAS REMOVED AND REPLACED WITH A NEW NAIL AND HELICAL BLADE WITH NO FURTHER PROBLEMS. IT WAS NOTED THAT ALL PARTS OF THE ORIGINAL NAIL WERE INTACT AND THEREFORE, THERE WERE NO FRAGMENTS TO RETRIEVE. THE PROCEDURE WAS EXTENDED BY APPROXIMATELY 30 MINUTES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
29612 12MM/130 DEG TI CANN TROCH FIXATION NAIL 380MM/RIGHT-STER TROCH FIXATION NAIL HSB SYNTHES MONUMENT 7015360

Patients

Seq Age Sex Outcome Treatment
1 66 YR Required Intervention HELICAL BLADE