OT ULTRAMINI METER
Report
- Report Number
- 3008382007-2013-01116
- Event Type
- Injury
- Date Received
- January 22, 2013
- Date of Event
- January 9, 2013
- Report Date
- January 13, 2013
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K061118
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP ((B)(4) 2013)-DEVICE EVALUATION: THE METER INVOLVED WITH THIS COMPLAINT HAS BEEN RETURNED AND EVALUATED BY LIFESCAN PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE REPORTED ISSUE COULD NOT BE REPRODUCED. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED. THE RETAIN TEST STRIPS WERE TESTED AND THEY PASSED TESTING WITH NO FAULTS FOUND. A DHR (DEVICE HISTORY RECORD) WAS COMPLETED FOR THIS PRODUCT AND NO DEVIATIONS, NON-CONFORMANCES, OR REWORKS WERE OBSERVED.
ON (B)(6) 2013, THE LAY USER/ PATIENT CONTACTED LIFESCAN (LFS) ALLEGING HER ONETOUCH ULTRAMINI METER READ INACCURATELY. THE COMPLAINT WAS CLASSIFIED BASED ON THE CUSTOMER CARE ADVOCATE (CCA) DOCUMENTATION. THE ALLEGED ISSUE BEGAN ON THE MORNING OF (B)(6) 2013. THE PATIENT REPORTED A BLOOD GLUCOSE RESULT OF "147 MG/DL" WITH THE SUBJECT METER. THE PATIENT MANAGES HER DIABETES WITH INSULIN. IT IS NOT KNOWN IF THE PATIENT MADE CHANGES TO HIS USUAL DOSE OF MEDICATIONS. THE PATIENT DENIED DEVELOPING SYMPTOMS; HOWEVER, THE PATIENT REPORTEDLY OBTAINED A BLOOD GLUCOSE RESULT OF "23 MG/DL" WITH ANOTHER DEVICE. THE PATIENT DRANK JUICE AS TREATMENT. AT THE TIME OF TROUBLESHOOTING, THE CCA NOTED THE SUBJECT METER WAS SET TO THE CORRECT UNIT OF MEASUREMENT. THE PATIENT DID NOT HAVE CONTROL SOLUTION TO PERFORM QUALITY CONTROL TESTING. THIS COMPLAINT IS BEING REPORTED BECAUSE THE PATIENT OBTAINED A BLOOD GLUCOSE RESULT OF "23 MG/DL" WITH ANOTHER DEVICE WHICH IS SUGGESTIVE OF A SERIOUS INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 30541 | OT ULTRAMINI METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3329242 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | Life Threatening| R |