FDA Adverse Event Other Summary report: N

LIBERTY 624EZ

MDR report key: 2923339 · Received October 2, 2006

Report

Report Number
1065497-2006-00001
Event Type
Other
Date Received
October 2, 2006
Date of Event
September 11, 2006
Report Date
September 28, 2006
Manufacturer
EAGLE PRODUCTS
Product Code
ITI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON INITIAL INSPECTION IT WAS DETERMINED THE ORIGINAL MANUFACTURES BATTERY HARDWARE HAD BEEN CHANGED, WHEN THE BATTERIES WERE RECENTLY REPLACED BY DEALER. THE BOLTS THAT WERE USED FOR THE CONNECTION OF THE TERMINALS WERE TOO LONG AND INSTALLED BACKWARDS. THIS ALLOWED THE BATTERY POST TO COME INTO CONTACT WITH FRAME OF THE CHAIR. WHEN THIS OCCURRED THE BATTERY ARCED AND WELDED THE BATTERY POST HARDWARE TO THE FRAME OF THE CHAIR. THIS ALLOWED THE BATTERY TO OVERHEAT AND IGNITE CAUSING THE CHAIR TO CATCH ON FIRE. THIS CHAIR HAS BEEN IN USE 1 YEAR AND 2 MONTHS PRIOR TO THIS INCIDENT. THERE IS NO HISTORY OR EVIDENCE TO SUGGEST A PRODUCT PROBLEM OR MALFUNCTION. THE MANUFACTURER VIEWS THIS AS A DEALER/SERVICE ERROR.

Description of Event or Problem · 1

THE CUSTOMER WAS USING HIS POWER WHEELCHAIR OUTDOORS, WHEN HE SAW SPARKS COMING FROM UNDERNEATH HIS CHAIR. HE WAS ABLE TO GET OUT OF THE CHAIR AND AFTER, THE CHAIR CAUGHT ON FIRE AND BURNED COMPLETELY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIBERTY 624EZ WHEELCHAIR ITI EAGLE PRODUCTS 624EZ

Patients

Seq Age Sex Outcome Treatment
1 82 YR Other