FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2923176 · Received January 18, 2013

Report

Report Number
2032227-2013-00259
Event Type
Malfunction
Date Received
January 18, 2013
Date of Event
January 14, 2013
Report Date
January 14, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP HAD A FROZEN DISPLAY AND CONSTANT TONE DUE TO CORROSION ON THE MOTHER BOARD. THE INSULIN PUMP HAD A CRACKED DISPLAY WINDOW AND MISSING END CAP STICKER.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A CRACK ON THE CASE, NEAR THE RESERVOIR AND BATTERY COMPARTMENTS. THE CUSTOMER ALSO REPORTED FLUID INSIDE THE CASE DUE TO GOING INTO THE POOL WITH THE INSULIN PUMP ON. THE CUSTOMER STATED THAT SHE REMOVED THE BATTERY AND INSERTED A NEW ONE. THE CUSTOMER STATED THE INSULIN PUMP COUNTED UP TO THE NUMBER 6, THEN FROZE. PRIOR TO GOING INTO THE POOL, THE CUSTOMER HAD RECEIVED TWO ALARMS WITH A LOUD BEEP. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
27804 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-522LNAL

Patients

Seq Age Sex Outcome Treatment
1 28 YR