PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 2923166
·
Received January 18, 2013
Report
- Report Number
- 2032227-2013-00242
- Event Type
- Malfunction
- Date Received
- January 18, 2013
- Date of Event
- January 10, 2013
- Report Date
- January 10, 2013
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- PATIENT FAMILY MEMBER OR FRIEND
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP ALARMED DURING THE PRIME TEST DUE TO A LOOSE MOTOR SUPPORT DISK. THE INSULIN PUMP ALSO HAD A MISSING END CAP STICKER.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED DURING THE MANUAL PRIME. ADVISED THE CALLER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 27234 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-522LNAH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 16 YR |