FDA Adverse Event Malfunction Summary report: N

TRANSTAR STRETCHER

MDR report key: 2923090 · Received December 27, 2012

Report

Report Number
1824206-2012-08627
Event Type
Malfunction
Date Received
December 27, 2012
Date of Event
December 10, 2012
Report Date
December 10, 2012
Manufacturer
HILL-ROM, INC.
Product Code
FPO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECH ADJUSTED BRAKE CASTERS TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

THE TECH REPORTED THAT THE BRAKE CASTERS ARE NOT HOLDING. NO INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TRANSTAR STRETCHER WHEELED STRETCHER FPO HILL-ROM, INC. 8000

Patients

Seq Age Sex Outcome Treatment
1