FDA Adverse Event
Malfunction
Summary report: N
SPECTRUM INFUSION PUMP
MDR report key: 2923031
·
Received December 27, 2012
Report
- Report Number
- 1314492-2012-00535
- Event Type
- Malfunction
- Date Received
- December 27, 2012
- Date of Event
- November 19, 2012
- Report Date
- November 30, 2012
- Manufacturer
- BAXTER HEALTHCARE CORP
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). SIGMA EVALUATED THE DEVICE IN QUESTION. IT WAS CONFIRMED THAT THE AUDIO FUNCTION WAS NOT PRESENT. FURTHER EVAL IDENTIFIED THAT THE ABSENCE OF AUDIO WAS DUE TO A FAILED 1 WATT SPEAKER. ADDITIONALLY, THE PUMP WAS TESTED AT MULTIPLE RATES AND NO ANOMALOUS NOISES WERE OBSERVED.
Description of Event or Problem · 1
IT WAS REPORTED THAT A PUMP MOTOR WAS "CLICKING" AND THE SPEAKER IS INOPERABLE. IT WAS ALSO REPORTED THAT THERE WAS NO PT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPECTRUM INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE CORP | 35700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Congenital Anomaly |