FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 2923031 · Received December 27, 2012

Report

Report Number
1314492-2012-00535
Event Type
Malfunction
Date Received
December 27, 2012
Date of Event
November 19, 2012
Report Date
November 30, 2012
Manufacturer
BAXTER HEALTHCARE CORP
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). SIGMA EVALUATED THE DEVICE IN QUESTION. IT WAS CONFIRMED THAT THE AUDIO FUNCTION WAS NOT PRESENT. FURTHER EVAL IDENTIFIED THAT THE ABSENCE OF AUDIO WAS DUE TO A FAILED 1 WATT SPEAKER. ADDITIONALLY, THE PUMP WAS TESTED AT MULTIPLE RATES AND NO ANOMALOUS NOISES WERE OBSERVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PUMP MOTOR WAS "CLICKING" AND THE SPEAKER IS INOPERABLE. IT WAS ALSO REPORTED THAT THERE WAS NO PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPECTRUM INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE CORP 35700

Patients

Seq Age Sex Outcome Treatment
1 Congenital Anomaly