FDA Adverse Event Malfunction Summary report: N

AFFINITY THREE BIRTHING BED

MDR report key: 2923024 · Received December 27, 2012

Report

Report Number
1824206-2012-08589
Event Type
Malfunction
Date Received
December 27, 2012
Date of Event
December 11, 2012
Report Date
December 11, 2012
Manufacturer
HILL-ROM INC.
Product Code
HDD
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THAT THE TREND LINKAGE WAS NOT ADJUSTED PROPERLY. THE BED WOULD NOT STAY IN TREND HANDLES WERE ENGAGED BECAUSE THE CYLINDER DID NOT LOCK INTO POSITION. THE TECHNICIAN ADJUSTED THE TREND LINKAGE TO REPAIR THE BED.

Description of Event or Problem · 1

INFO REC'D INDICATES THE BED WILL NOT STAY IN THE TREND POSITION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AFFINITY THREE BIRTHING BED BIRTHING BED HDD HILL-ROM INC. 3700

Patients

Seq Age Sex Outcome Treatment
1