ARTICULEZE M HEAD 36MM +12
Report
- Report Number
- 1818910-2013-22350
- Event Type
- Injury
- Date Received
- January 18, 2013
- Report Date
- December 21, 2012
- Manufacturer
- DEPUY ORTHOPAEDICS, INC. 1818910
- Product Code
- JDI
- PMA / PMN Number
- K980513
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- ATTORNEY
Narratives
THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.
THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. A SEARCH OF THE COMPLAINT DATABASE FOUND NO ADDITIONAL REPORTS FOR THE LOT CODE 2021903. A SEARCH OF THE COMPLAINT DATABASE SEARCH FINDS ONE ADDITIONAL REPORTED INCIDENT AGAINST LOT CODE 2023708 SINCE ITS RELEASE FOR DISTRIBUTION; HOWEVER, A PREVIOUS REVIEW OF THE DEVICE HISTORY RECORD DID NOT REVEAL ANY RELATED MANUFACTURING ANOMALIES. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO DETERMINE A ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD ANY ADDITIONAL INFORMATION BE RECEIVED TO CHANGE THE OUTCOME OF THE PERFORMED INVESTIGATION, THE COMPLAINT WILL BE RE-OPENED.
NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. BASED ON PREVIOUS INVESTIGATIONS THIS COMPLICATION OF JOINT REPLACEMENT IS UNLIKELY TO HAVE BEEN THE RESULT OF A DEVICE FAILING TO MEET REQUIRED SPECIFICATIONS. THE INFORMATION RECEIVED WILL BE RETAINED FOR POTENTIAL SERIES INVESTIGATIONS IF TRIGGERED BY TREND ANALYSIS, POST MARKET SURVEILLANCE, OR OTHER EVENTS WITHIN THE QUALITY SYSTEM. CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.
(B)(4). INVESTIGATION SUMMARY: NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. BASED ON PREVIOUS INVESTIGATIONS THIS COMPLICATION OF JOINT REPLACEMENT IS UNLIKELY TO HAVE BEEN THE RESULT OF A DEVICE FAILING TO MEET REQUIRED SPECIFICATIONS. THE INFORMATION RECEIVED WILL BE RETAINED FOR POTENTIAL SERIES INVESTIGATIONS IF TRIGGERED BY TREND ANALYSIS, POST MARKET SURVEILLANCE, OR OTHER EVENTS WITHIN THE QUALITY SYSTEM. CORRECTIVE ACTION WAS NOT INDICATED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.
(B)(4).
LITIGATION PAPERS ALLEGE THE FOLLOWING: AFTER THE SURGERY, FRICTION AND WEAR BETWEEN THE COBALT-CHROMIUM METAL HEAD AND COBALT-CHROMIUM METAL LINER CAUSED LARGE AMOUNTS OF TOXIC COBALT-CHROMIUM METAL IONS AND PARTICLES TO BE RELEASED INTO PATIENTS BLOOD AND TISSUE AND BONE SURROUNDING THE IMPLANT. AS A RESULT, PATIENT HAS BEEN EXPERIENCING SEVERE PAIN AND DISCOMFORT AND INFLAMMATION IN HIS RIGHT THIGH AND GROIN. HE ALSO EXPERIENCES A POPPING AND SNAPPING SENSATION IN HIS HIP-JOINT WHEN WALKING OR MOVING TO AND FROM A SITTING POSITION. DUE TO PATIENTS CHRONIC PAIN AND DISCOMFORT AND OTHER SYMPTOMS, PATIENT WILL LIKELY NEED TO UNDERGO REVISION SURGERY TO REPLACE THE IMPLANT. **UPDATE**(B)(4) 2012 - PFS AND MEDICAL RECORDS WERE RECEIVED FROM LEGAL. PART/LOT INFORMATION WAS IDENTIFIED. RECORDS ARE AVAILABLE FOR FURTHER REVIEW.
THERE WERE NO NEW ALLEGATION. ADDED LAWYER, LAW FIRM AND UNKNOWN STEM. DOI: (B)(6) 2006; DOR: (B)(6) 2015; (RIGHT HIP).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 28287 | ARTICULEZE M HEAD 36MM +12 | ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS | JDI | DEPUY ORTHOPAEDICS, INC. 1818910 | 2023708 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 41 YR | Other |