FDA Adverse Event Malfunction Summary report: N

POSEY KEEPSAFE DELUXE

MDR report key: 2922757 · Received December 28, 2012

Report

Report Number
2020362-2012-00698
Event Type
Malfunction
Date Received
December 28, 2012
Report Date
December 4, 2012
Manufacturer
J. T. POSEY CO.
Product Code
KMI
PMA / PMN Number
EXEMPT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

RESULTS: EVAL OF THE RETURNED PRODUCT CONFIRMED THE REPORTED ISSUE THAT THE ALARM DOES NOT POWER ON WITH BATTERIES. THE ALARM DOES POWER UP WHEN USING AN EXTERNAL POWER SUPPLY AND/OR A 9V ADAPTER. THE VOICE MESSAGE PLAYS BUT WITH STATIC. THE ALARM PASSES ALL OTHER FUNCTIONAL TESTS. A FLAT BATTERY CONTACT IS CORRODED DUE TO BATTERY LEAKAGE. NOTE: INSTRUCTIONS FOR USE HAS A WARNING STATEMENT ON BATTERY INSTALLATION: TAKE CARE WHEN INSTALLING NEW BATTERIES. THE ALARM WILL NOT WORK IF BATTERIES ARE INSTALLED IMPROPERLY. ALWAYS INSTALL A COMPLETELY NEW SET OF BATTERIES WHEN THE CHIRPING DUE TO LOW BATTERY POWER SOUND BEGINS. DO NOT REPLACE A SINGLE CELL, BUT ALL CELLS IN THE ALARM. DO NOT MIX OLD AND NEW BATTERIES, OR BATTERY BRANDS WITHIN A BATTERY PACK (4 BATTERIES). USE OF MIXED BATTERIES OR BATTERIES INSTALLED INCORRECTLY MAY CAUSE BATTERY DAMAGE, AND MAY DAMAGE THE ALARM. REMOVE ANY ALARM FROM USE AND SEND TO THE APPROPRIATE FACILITY AUTHORITY IF BATTERIES ARE DAMAGED OR CORRODED OR THE BATTERY COMPARTMENT HAS SIGNS OF PREVIOUS BATTERY CORROSION SUCH AS WHITE POWDER RESIDUE. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE ALARM WILL NOT POWER ON EVEN AFTER THE BATTERIES WERE REPLACED WITH NEW ONES. THE CUSTOMER REPORTED THAT THERE ARE NO SIGNS OF BATTERY LEAKAGE OR CORROSION IN THE BATTERY COMPARTMENT. THE CUSTOMER DID NOT PROVIDE A DATE WHEN THIS OCCURRED. THERE WAS NO PT INCIDENT OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 POSEY KEEPSAFE DELUXE KMI J. T. POSEY CO. 8374 NA

Patients

Seq Age Sex Outcome Treatment
1 UNK SQUARE CHAIR SENSOR PAD: MODEL 8308, LOT # UNK