FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2922755 · Received January 18, 2013

Report

Report Number
1416980-2013-01401
Event Type
Malfunction
Date Received
January 18, 2013
Date of Event
December 1, 2012
Report Date
December 28, 2012
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915523
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INITIAL EVALUATION WAS PERFORMED ON-SITE BY A BAXTER FIELD SERVICE TECHNICIAN AND CONFIRMED THE REPORTED CONDITION. UPON COMPLETION OF BAXTER'S INVESTIGATION, A FOLLOW-UP WILL BE SUBMITTED. A SERVICE HISTORY REVIEW WAS PERFORMED REVEALING THE REPORTED CONDITION IS NOT RELATED TO ANY PREVIOUS CUSTOMER SERVICE REQUESTS ON THIS PUMP.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF "F-38 ALARM" WAS CONFIRMED DURING DEVICE EVALUATION. THE ROOT CAUSE WAS DUE TO DAMAGED FORCE SENSING RESISTORS (FSRS). THE FSRS WERE REPLACED TO RESOLVE THE REPORTED CONDITION.

Description of Event or Problem · 1

BAXTER (B)(4) REPORTED A FLOGARD INFUSION PUMP WITH A F-38 ALARM. IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT; THEREFORE, NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
27237 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1