FDA Adverse Event
Malfunction
Summary report: N
SPECTRUM INFUSION PUMP
MDR report key: 2922719
·
Received December 28, 2012
Report
- Report Number
- 1314492-2012-00553
- Event Type
- Malfunction
- Date Received
- December 28, 2012
- Report Date
- December 5, 2012
- Manufacturer
- BAXTER HEALTHCARE CORPORATION
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Removal / Correction Number
- 1314491-080712-002-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE SIGMA DEVICE EVAL FOUND THE #2, #3, #4, #5, #6, #7, #8, #9 AND THE (.) KEYS TO BE INOPERABLE. IT WAS OBSERVED THAT WHEN ANY REMAINING FUNCTIONAL KEYS ARE SELECTED, AN AUTOMATIC OUTPUT OF THE #3 KEY WILL OCCUR FOLLOWING THE SELECTED KEY'S FUNCTION (E.G. WHEN THE #1 KEY IS PRESSED 13 WILL BE DISPLAYED. WHEN IN ALPHABETIC MODE AND THE ABC KEY IS SELECTED, AG WILL BE DISPLAYED INTERFERING WITH THE MDL DRUG SEARCH). SIGMA'S ENGINEERING INVESTIGATION IS IN PROGRESS. WHEN COMPLETE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DATE OF EVENT IS UNK.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE "G" KEY ON A PUMP KEYPAD IS "STUCK." IT WAS ALSO REPORTED THAT THERE WAS NO PT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPECTRUM INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE CORPORATION | 35700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |