FDA Adverse Event
Malfunction
Summary report: N
SPECTRUM INFUSION UMP
MDR report key: 2922673
·
Received December 27, 2012
Report
- Report Number
- 1314492-2012-00542
- Event Type
- Malfunction
- Date Received
- December 27, 2012
- Date of Event
- November 26, 2012
- Report Date
- December 3, 2012
- Manufacturer
- BAXTER HEALTHCARE CORP
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE SIGMA DEVICE EVAL FOUND THE #2, #3, #4, #5, #6, #7, #8, #9 AND (.) KEYS TO BE INOPERABLE. IT WAS OBSERVED THAT WHEN ANY REMAINING FUNCTIONAL KEYS ARE SELECTED, AN AUTOMATIC OUTPUT OF THE #3 KEY WILL OCCUR FOLLOWING THE SELECTED KEY¿S FUNCTION (E.G. WHEN THE #1 KEY IS PRESSED 13 WILL BE DISPLAYED. WHEN IN ALPHABETIC MODE AND THE ¿ABC¿ KEY IS SELECTED, AG WILL BE DISPLAYED INTERFERING WITH THE MDL DRUG SEARCH). SIGMA¿S ENGINEERING INVESTIGATION IS IN PROGRESS. WHEN COMPLETE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE #3 KEY ON A PUMP KEYPAD IS "STUCK." THE CUSTOMER ALSO STATED THAT THERE WAS NO PATIENT INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPECTRUM INFUSION UMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE CORP | 35700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |