FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION UMP

MDR report key: 2922673 · Received December 27, 2012

Report

Report Number
1314492-2012-00542
Event Type
Malfunction
Date Received
December 27, 2012
Date of Event
November 26, 2012
Report Date
December 3, 2012
Manufacturer
BAXTER HEALTHCARE CORP
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE SIGMA DEVICE EVAL FOUND THE #2, #3, #4, #5, #6, #7, #8, #9 AND (.) KEYS TO BE INOPERABLE. IT WAS OBSERVED THAT WHEN ANY REMAINING FUNCTIONAL KEYS ARE SELECTED, AN AUTOMATIC OUTPUT OF THE #3 KEY WILL OCCUR FOLLOWING THE SELECTED KEY¿S FUNCTION (E.G. WHEN THE #1 KEY IS PRESSED 13 WILL BE DISPLAYED. WHEN IN ALPHABETIC MODE AND THE ¿ABC¿ KEY IS SELECTED, AG WILL BE DISPLAYED INTERFERING WITH THE MDL DRUG SEARCH). SIGMA¿S ENGINEERING INVESTIGATION IS IN PROGRESS. WHEN COMPLETE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE #3 KEY ON A PUMP KEYPAD IS "STUCK." THE CUSTOMER ALSO STATED THAT THERE WAS NO PATIENT INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SPECTRUM INFUSION UMP INFUSION PUMP FRN BAXTER HEALTHCARE CORP 35700

Patients

Seq Age Sex Outcome Treatment
1