FDA Adverse Event Malfunction Summary report: N

X-STREAM

MDR report key: 2922668 · Received December 27, 2012

Report

Report Number
1213643-2012-00902
Event Type
Malfunction
Date Received
December 27, 2012
Report Date
December 3, 2012
Manufacturer
DAVOL INC., SUB. C.R. BARD, INC.
Product Code
GCJ
PMA / PMN Number
K003790
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE RETURNED UNITS WERE FOUND TO HAVE CRACKED BLISTER TRAYS. THE BLISTER PACKAGING SEALS WERE NOTED TO HAVE BEEN FULLY FORMED AT ONE TIME. OUR INVESTIGATION DETERMINED THAT IT IS POSSIBLE THAT AN EVENT OF THIS NATURE COULD OCCUR IF THE DEVICES RECEIVED SOME SIGNIFICANT IMPACT DURING THE SHIPPING PROCESS. THE DAMAGE THAT CAN OCCUR TO THE BLISTERS IS A CLEAR BREACH OF THE STERILE BARRIER; HOWEVER, THIS DAMAGE IS HIGHLY EVIDENT TO THE USER. THE IFU CAUTIONS THAT PRODUCT IS STERILE UNLESS PACKAGE IS DAMAGED OR OPEN.

Description of Event or Problem · 1

BASED ON INFORMATION REPORTED TO DAVOL: IT WAS REPORTED THAT DURING SET UP FOR A CASE, X-STREAM IRRIGATION UNITS FROM INVENTORY WERE FOUND WITH CRACKED TRAYS. FURTHER REVIEW OF INVENTORY FOUND FIVE ADDITIONAL IRRIGATION UNITS TO BE IN THIS CONDITION. THESE IRRIGATION KITS CAN NOT BE USED AS THE CRACKED PACKAGING CAUSES LOSS OF STERILITY TO THE PRODUCT INSIDE. THERE WAS NO PATIENT INTERVENTION. DUE TO THE REPORTED STERILITY BREACH, THIS MDR IS BEING SUBMITTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 X-STREAM GCJ DAVOL INC., SUB. C.R. BARD, INC. NA JUWEF023

Patients

Seq Age Sex Outcome Treatment
1 NA