FDA Adverse Event Malfunction Summary report: N

ADVANTA 2

MDR report key: 2922638 · Received December 26, 2012

Report

Report Number
3006697241-2012-00312
Event Type
Malfunction
Date Received
December 26, 2012
Date of Event
December 6, 2012
Report Date
December 6, 2012
Manufacturer
HILLROM DE MEXICO S DE RL DE CV
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE BUTTONS ON CONTROL PANEL WAS PUSHED IN, ALLOWING IT TO MAKE CONTACT WHEN THE BED WAS PLUGGED IN. THE TECHNICIAN REPLACED BOTH THE LEFT AND RIGHT CAREGIVER CONTROL PANEL TO REPAIR THE BED.

Description of Event or Problem · 1

INFORMATION RECEIVED INDICATES THE BED IS RAISING AND LOWERING UNINTENTIONALLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADVANTA 2 AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILLROM DE MEXICO S DE RL DE CV 1190

Patients

Seq Age Sex Outcome Treatment
1