FDA Adverse Event Injury Summary report: N

ANEURX ABDOMINAL STENT GRAFT SYSTEM

MDR report key: 2922581 · Received January 18, 2013

Report

Report Number
2953200-2013-00117
Event Type
Injury
Date Received
January 18, 2013
Date of Event
December 21, 2012
Report Date
January 30, 2013
Manufacturer
MEDTRONIC CARDIOVASCULAR SANTA ROSA
Product Code
MIH
PMA / PMN Number
P990020
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION RESULTS: INHERENT RISK OF PROCEDURE (STENT GRAFT MIGRATION, ENDOLEAK, ANEURYSM RUPTURE). PATIENT'S CONDITION AFFECTED EFFECTIVENESS OF DEVICE (ANGULATED AORTIC NECK). CONCLUSION: DEVICE FAILURE/LACK OF EFFECTIVENESS RELATED TO PATIENT CONDITION (ANGULATED AORTIC NECK).

Description of Event or Problem · 1

AN ANEURX STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM APPROXIMATELY 52 MONTHS AGO. ANEURYSM MORPHOLOGY FROM THE TIME OF IMPLANT IS UNKNOWN. THE PATIENT HAS AN ANGULATED AORTIC NECK. AN ANEURX BIFURCATED STENT GRAFT BFXCH2414165, AN AORTIC CUFF AEXCH282840, AN IEXCH182485 AND AN ILXCH1616135 WERE IMPLANTED AND THERE WERE NO COMPLICATION SINCE THE TIME OF IMPLANT. IT WAS REPORTED THAT APPROXIMATELY FOUR WEEKS AGO THE PATIENT PRESENT EMERGENTLY WITH PAIN AND A CT SCAN WAS DONE WHICH REVEALED A RUPTURE. THE BIFURCATED STENT GRAFT WAS FOUND TO HAVE MIGRATED 4 CM WITH A PROXIMAL TYPE I ENDOLEAK PRESENT. AN ENDURANT AORTIC CUFF ETCF2828C49E AND AN ETEW2828C82E WERE IMPLANTED TO RESOLVE PROXIMAL TYPE I ENDOLEAK AND ANEURYSM RUPTURE. THE POSSIBLE CAUSE OF THIS EVENT WAS LIKELY DUE TO THE STENT GRAFT MIGRATION. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED AND THE PATIENT IS FINE.

Description of Event or Problem · 1

ADDITIONAL INFORMATINO WAS RECEIVED AS FOLLOWS: THE PHYSICIAN ELECTED TO IMPLANT ILIAC LIMBS BILATERALLY AS A PROPHYLACTIC MEASURE. THE CAUSE OF THE STENT GRAFT MIGRATION, TYPE I ENDOLEAK WHICH RESULTED IN ANEURYSM RUPTURE WAS RELATED TO THE SEVERE AORTIC NECK ANGULATION. IT IS UNKNOWN IF THE AORTIC NECK ANGULATION WAS AT THE TIME OF IMPLANT OR IF IT WAS DISEASE PROGRESSION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
27947 ANEURX ABDOMINAL STENT GRAFT SYSTEM SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT MIH MEDTRONIC CARDIOVASCULAR SANTA ROSA V00027009Y

Patients

Seq Age Sex Outcome Treatment
1 00084 YR Required Intervention