PROGRAMMING SOFTWARE
Report
- Report Number
- 1644487-2013-00191
- Event Type
- Malfunction
- Date Received
- January 18, 2013
- Date of Event
- September 1, 2012
- Report Date
- December 21, 2012
- Manufacturer
- CYBERONICS, INC.
- Product Code
- LYJ
- PMA / PMN Number
- P970003
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE FAILURE OCCURRED, BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY.
DEVICE FAILURE IS SUSPECTED, BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY.
ANALYSIS OF THE VNS COMPUTER AND WAND WAS COMPLETED. DURING THE ANALYSIS IT WAS IDENTIFIED THAT THE HANDHELD COMPUTER WAS UNABLE TO ADVANCE PAST THE SCREEN ALIGNMENT UTILITY. THE CAUSE FOR THE DISPLAY ANOMALY IS ASSOCIATED WITH RESISTANCE VALUES BEING HIGHER THAN EXPECTED IN THE TOUCH SCREEN CIRCUITRY. CONTINUITY TESTING OF THE RETURNED DISPLAY IDENTIFIED AN ANOMALY ASSOCIATED WITH THE TRACE THAT IS ROUTED TO PIN 4 OF THE TOUCH SCREEN FLEX CABLE. RESISTANCE READINGS ASSOCIATED WITH THE CIRCUIT WERE HIGHER THAN EXPECTED (READ APPROXIMATELY 57M OHMS FOR PIN 4 - NOMINAL IS APPROXIMATELY .418K OHMS). PRESSED ON THE FLEX CABLE WHERE IT ATTACHES TO THE TOUCH SCREEN AND IT WAS IDENTIFIED THAT THE RESISTANCE READING WAS STILL SIGNIFICANTLY HIGHER THAN EXPECTED. ONCE THE DISPLAY WAS REPLACED WITH A KNOWN GOOD DISPLAY, NO FURTHER ANOMALIES ASSOCIATED WITH THE HANDHELD COMPUTER PERFORMANCE WERE IDENTIFIED DURING THE ANALYSIS. NO ANOMALIES ASSOCIATED WITH FLASHCARD SOFTWARE OR DATABASES WERE IDENTIFIED DURING THE FLASHCARD ANALYSIS. THE FLASHCARD AND SOFTWARE PERFORMED ACCORDING TO FUNCTIONAL SPECIFICATIONS
REPORTER INDICATED VNS PROGRAMMING SYSTEM WAS HAVING INTERMITTENT COMMUNICATION ISSUES SINCE (B)(6) 2012, AND NOW WAS NOT COMMUNICATING AT ALL. IT WAS LATER REPORTED THAT THE COMPUTER SCREEN WOULD NOT ADVANCE PAST THE ALIGN SCREEN SINCE THE LATEST CYBERONICS SOFTWARE UPGRADE FROM VERSION (B)(4). CLEANING THE SCREEN DID NOT RESOLVE THE ISSUE AND THE SCREEN WOULD STILL NOT ADVANCE PAST THE ALIGN SCREEN. THE PROGRAMMING WAND WAS CONFIRMED TO BE PERFORMING AS INTENDED WITH A DIFFERENT VNS COMPUTER. THE COMPUTER AND FLASHCARD WERE RECEIVED ON 01/09/2013 AND ANALYSIS IS PENDING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 28042 | PROGRAMMING SOFTWARE | PROGRAMMING COMPUTER | LYJ | CYBERONICS, INC. | MODEL 250 | 1075844 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |