FDA Adverse Event Malfunction Summary report: N

PROGRAMMING SOFTWARE

MDR report key: 2922469 · Received January 18, 2013

Report

Report Number
1644487-2013-00190
Event Type
Malfunction
Date Received
January 18, 2013
Date of Event
September 1, 2012
Report Date
December 21, 2012
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE FAILURE IS SUSPECTED, BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY.

Description of Event or Problem · 1

ANALYSIS OF THE COMPUTER AND FLASHCARD WAS COMPLETED. NO ANOMALIES ASSOCIATED WITH THE HANDHELD COMPUTER WERE NOTED DURING TESTING USING THE AC ADAPTER OR THE MAIN BATTERY WITH A FULL CHARGE. THE HANDHELD COMPUTER PERFORMED ACCORDING TO FUNCTIONAL SPECIFICATIONS. NO ANOMALIES ASSOCIATED WITH FLASHCARD SOFTWARE OR DATABASES WERE IDENTIFIED DURING THE FLASHCARD ANALYSIS. THE FLASHCARD AND SOFTWARE PERFORMED ACCORDING TO FUNCTIONAL SPECIFICATIONS.

Description of Event or Problem · 1

REPORTER INDICATED A VNS PROGRAMMING SYSTEM WAS HAVING INTERMITTENT COMMUNICATION ISSUES SINCE (B)(6) 2012, AND WAS NOW NOT COMMUNICATING AT ALL. THE REPORTER WAS USING THE PROGRAMMING SYSTEM WHILE PLUGGED INTO THE WALL OUTLET, SO THIS MAY BE THE CAUSE. THE BATTERY IN THE PROGRAMMING WAND WAS TESTED AND WAS SUFFICIENT. THE COMPUTER, FLASHCARD, AND WAND WERE RETURNED FOR ANALYSIS ON (B)(4) 2013. NO ANOMALIES WERE IDENTIFIED DURING THE WAND ANALYSIS, AND THE WAND PERFORMED AS INTENDED. ANALYSIS OF THE COMPUTER AND FLASHCARD IS STILL PENDING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
28072 PROGRAMMING SOFTWARE PROGRAMMING SOFTWARE LYJ CYBERONICS, INC. MODEL 250 842377

Patients

Seq Age Sex Outcome Treatment
1