FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2922468 · Received January 18, 2013

Report

Report Number
1416980-2013-01382
Event Type
Malfunction
Date Received
January 18, 2013
Date of Event
December 1, 2012
Report Date
December 28, 2012
Manufacturer
SHARP CORPORATION OSAKA JAPAN
Product Code
FRN
PMA / PMN Number
K915522
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE IS CURRENTLY IN THE PROCESS OF BEING EVALUATED ON-SITE BY A BAXTER FIELD SERVICE TECHNICIAN. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4).EVALUATION SUMMARY:THE REPORTED CONDITION OF "BATTERY DAMAGED" WAS CONFIRMED BY QUALITY ENGINEERING AS A BATTERY LOW ALERT. THE ROOT CAUSE WAS DUE TO DEPLETED MAIN BATTERIES, WHICH WERE REPLACED TO RESOLVE THE REPORTED CONDITION. ADDITIONAL INFORMATION:A SERVICE HISTORY REVIEW WAS PERFORMED REVEALING THE REPORTED CONDITION IS NOT RELATED TO ANY PREVIOUS CUSTOMER SERVICE REQUEST ON THIS PUMP.

Description of Event or Problem · 1

BAXTER (B)(4) REPORTED A FLOGARD INFUSION PUMP WITH BATTERY DAMAGED. IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. THERE WAS NO REPORT OF PATIENT INVOLVEMENT; THEREFORE, NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
28228 FLOGARD PUMP, INFUSION FRN SHARP CORPORATION OSAKA JAPAN

Patients

Seq Age Sex Outcome Treatment
1