FDA Adverse Event Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2922167 · Received January 18, 2013

Report

Report Number
2531779-2013-00841
Date Received
January 18, 2013
Report Date
January 8, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 DATE OF SUBMISSION (B)(4) 2013 - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE PEELING AT THE OK BUTTON. A DAMAGED KEYPAD WILL ALLOW CONTAMINATION TO PERMEATE THE BUTTON CONTACTS NEGATIVELY IMPACTING THE BUTTON FUNCTION. THE DAMAGED KEYPAD IS OBVIOUS AND DETECTABLE AND SHOULD ALERT THE USER TO DISCONTINUE USE OF THE PUMP. DURING TESTING, THE DOWN ARROW AND OK KEYPAD BUTTONS WERE FOUND TO BE UNRESPONSIVE; THE UP ARROW AND CONTRAST BUTTONS HAD AN INTERMITTENT RESPONSE. DURING INVESTIGATION THE KEYPAD WAS REMOVED AND THE OK AND DOWN ARROW CONTACTS WERE FOUND TO BE MISSING. THERE WAS EVIDENCE OF CONTAMINATION FOUND UNDER ALL KEY CONTACTS. UNRELATED TO THE COMPLAINT, EVALUATION REVEALED A DIM/FADING DISPLAY SCREEN, WHICH HAS NO EFFECT ON INSULIN DELIVERY FUNCTION.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

THE REPORTER CONTACTED ANIMAS ON (B)(4) 2013 STATING THAT THE DOWN ARROW KEYPAD BUTTON WAS UNRESPONSIVE AFTER HAVING BEEN INTERMITTENTLY UNRESPONSIVE. THE REPORTER NOTED DAMAGE TO THE KEYPAD AND WAS UNABLE TO DETERMINE IF THE DAMAGE OR THE RESPONSE ISSUE HAD OCCURRED FIRST. THERE IS NO PATIENT IMPACT WITH THIS EVENT. THIS COMPLAINT IS BEING REPORTED DUE TO THE ALLEGED KEYPAD ISSUES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
27482 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 18 YR