FDA Adverse Event Other Summary report: N

*

MDR report key: 292205 · Received August 11, 2000

Report

Report Number
9611451-2000-00003
Event Type
Other
Date Received
August 11, 2000
Date of Event
June 23, 2000
Report Date
August 11, 2000
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BTT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE HEATER WIRE OF THE DUAL HEATED ALLEGIANCE BREATHING CIRCUIT (#7100-4S2) HAD MELTED THROUGH THE WALL OF THE INSPIRATORY LIMB. A SINGLE LIMB ALLEGIANCE ELECTRICAL ADAPTOR (#0060-20) HAD BEEN USED WITH THE CIRCUIT. THIS LED TO THE INSPIRATORY HEATER WIRE OVERHEATING. THERE WERE NO INJURIES TO THE NEONATAL PT. INSERVICE TRAINING OF THE STAFF AS TO THE PROPER SET-UP OF THE EQUIPMENT HAS BEEN AGREED TO.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 * HEATED RESPIRATORY HUMIDIFIER BTT FISHER & PAYKEL HEALTHCARE LTD MR730JHY *

Patients

Seq Age Sex Outcome Treatment
1 *