FDA Adverse Event
Other
Summary report: N
*
MDR report key: 292205
·
Received August 11, 2000
Report
- Report Number
- 9611451-2000-00003
- Event Type
- Other
- Date Received
- August 11, 2000
- Date of Event
- June 23, 2000
- Report Date
- August 11, 2000
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THE HEATER WIRE OF THE DUAL HEATED ALLEGIANCE BREATHING CIRCUIT (#7100-4S2) HAD MELTED THROUGH THE WALL OF THE INSPIRATORY LIMB. A SINGLE LIMB ALLEGIANCE ELECTRICAL ADAPTOR (#0060-20) HAD BEEN USED WITH THE CIRCUIT. THIS LED TO THE INSPIRATORY HEATER WIRE OVERHEATING. THERE WERE NO INJURIES TO THE NEONATAL PT. INSERVICE TRAINING OF THE STAFF AS TO THE PROPER SET-UP OF THE EQUIPMENT HAS BEEN AGREED TO.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | * | HEATED RESPIRATORY HUMIDIFIER | BTT | FISHER & PAYKEL HEALTHCARE LTD | MR730JHY | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |