FDA Adverse Event Malfunction Summary report: N

SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE INTRAOCUL

MDR report key: 2922033 · Received January 15, 2013

Report

Report Number
2023826-2013-00041
Event Type
Malfunction
Date Received
January 15, 2013
Date of Event
December 17, 2012
Report Date
December 19, 2012
Manufacturer
STAAR SURGICAL CO.
Product Code
HQL
PMA / PMN Number
P880091
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE INTRAOCULAR LENS (ELASTIC) WITH TORIC OPTIC. (B)(4). CONCLUSION - AN INVESTIGATION OF THE ROOT CAUSES OF LENS TEARS, SIMILAR TO THAT STATED IN THIS CLAIM, WERE ADDRESSED IN A CAPA OPENED IN JUNE 2005 (WHICH WAS SUBSEQUENTLY CLOSED). THE CAPA ADDRESSED DELIVERY SYSTEM ISSUES INCLUDING ALL STAGES OF MANUFACTURING OF THE INJECTORS AND CARTRIDGES. PROCESSES WERE REVIEWED AND REVISED AS OPPORTUNITIES FOR IMPROVEMENT WERE REVEALED. THE CAPA ALSO ADDRESSED HANDLING ERRORS. ALL INJECTOR/CARTRIDGE DIRECTIONS FOR USE (DFU) HAVE BEEN MODIFIED TO ADD FURTHER CLARIFICATIONS TO INSTRUCT THE USERS IN THE PROPER DELIVERY TECHNIQUES THAT ARE EFFECTIVE, AND MINIMIZE THE POTENTIAL FOR DAMAGING THE LENS. THE DAMAGE TO THE LENS COULD NOT BE EVALUATED BECAUSE LENS WAS NOT RETURNED TO STAAR. BASED ON THE COMPLAINT HISTORY, POSSIBLE ROOT CAUSES FOR LENS TEARS INCLUDE BOTH DELIVERY SYSTEM ISSUES AND HANDLING ERRORS BY THE CUSTOMER. (B)(4).

Description of Event or Problem · 1

THE REPORTER STATED THE SURGEON INSERTED AN AA4203TL TORIC OPTIC SILICONE SINGLE PIECE LENS. STATED THE LENS TORE DURING INSERTION INTO THE PT'S EYE. THE LENS WAS CUT AND TAKEN OUT OF EYE IN PIECES. UNABLE TO SAVE THE LENS. LENS WAS DISCARDED. THERE WAS NO PT INJURY. A BACKUP LENS, SAME MODEL AND SIZE WAS IMPLANTED. NO LOT NUMBERS WERE PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
22559 SILICONE ULTRAVIOLET-ABSORBING POSTERIOR CHAMBER SINGLE PIECE FOLDABLE INTRAOCUL INTRAOCULAR LENS HQL STAAR SURGICAL CO. AA4203TL NA

Patients

Seq Age Sex Outcome Treatment
1 CARTRIDGE: MODEL: MTC-60C FP, LOT NUMBER UNK| INJECTOR: MODEL: MSI-PR, LOT NUMBER UNK