FDA Adverse Event Injury Summary report: N

ACIST

MDR report key: 2921936 · Received January 11, 2013

Report

Report Number
2134243-2012-00020
Event Type
Injury
Date Received
January 11, 2013
Date of Event
November 12, 2012
Report Date
January 11, 2013
Manufacturer
ACIST MEDICAL SYSTEMS, INC.
Product Code
DXT
PMA / PMN Number
K991103
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE ACIST CONTRAST INJECTION SYSTEM, MODEL CVI, WAS TESTED ON (B)(4) 2012. THE INJECTION SYSTEM WAS FUNCTIONALLY TESTED AND THERE WAS NO EVIDENCE OF DEVICE MALFUNCTION. THE CONSUMABLE KITS USED DURING THE EVENT WERE DISCARDED BY THE HOSPITAL; THEREFORE, NO ANALYSIS COULD BE MADE OF THESE ITEMS. THE ACIST CONTRAST INJECTION SYSTEM IS EQUIPPED WITH AN AIR COLUMN DETECT SENSOR. THE AIR COLUMN DETECT SENSOR SENSES AIR IN THE PROXIMAL END OF THE HIGH-PRESSURE (INJECTION) TUBING. IF AIR IS DETECTED IN THE TUBING, ALL FLUID DELIVERY FUNCTIONS ARE DISABLED. PER THE ACIST CVI USERS MANUAL, THE AIR COLUMN DETECT SENSOR IS DESIGNED TO AID THE USER IN THE DETECTION OF AIR COLUMNS IN THE INJECTION LINE, BUT IT IS NOT DESIGNED TO REPLACE THE VIGILANCE AND CARE REQUIRED OF THE OPERATOR IN VISUALLY INSPECTING FOR AIR AND CLEANING AIR FROM THE ENTIRE PT KIT AND ANGIOGRAPHIC CATHETER. THE AIR COLUMN DETECT MECHANISM IS TO BE USED IN CONJUNCTION WITH AND TO COMPLEMENT THE USER'S OTHER PROCEDURES FOR PREVENTING AIR INJECTIONS. ACIST'S RISK MGMT HAS APPROPRIATE RISK MITIGATIONS IN PLACE FOR POTENTIAL AIR EMBOLISM DUE TO USER ERROR, INCLUDING LABELING DESCRIBING AIR BUBBLE PRECAUTIONS AND THE USER MANUAL WHICH GUIDES THE USER THROUGH SET-UP AND PURGE OF THE INJECTOR SYSTEM. BASED ON TESTING OF THE INJECTOR, THERE IS NO EVIDENCE OF DEVICE MALFUNCTION. A F/U REPORT WILL BE SUBMITTED UPON COMPLETION OF REVIEW OF THE CINE-ANGIOGRAM BY ACIST'S MEDICAL ADVISORY BOARD.

Description of Event or Problem · 1

DURING A DIAGNOSTIC CATHETERIZATION OF THE RCA, AIR WAS INJECTED INTO THE PT. THE PT HAD CHEST PAIN AND ST SEGMENT EVAL FOR APPROX 5 MINS DURATION, HYPOTENSION, AND ASYSTOLE. THE PT WAS TREATED WITH CPR AND RECOVERED. THE PHYSICIAN PROCEEDED WITH PERCUTANEOUS CORONARY INTERVENTION. (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
19246 ACIST ANGIOGRAPHIC INJECTION SYSTEM DXT ACIST MEDICAL SYSTEMS, INC. CVI NI

Patients

Seq Age Sex Outcome Treatment
1 66 YR Life Threatening| R CLOPIDOGREL (CLOPIDOGREL)| ASA (ACETYLSALICYLIC ACID)