FDA Adverse Event Malfunction Summary report: N

GEMSTAR DOCKING STATION

MDR report key: 2921846 · Received January 15, 2013

Report

Report Number
3008785455-2013-00008
Event Type
Malfunction
Date Received
January 15, 2013
Date of Event
August 1, 2012
Report Date
December 20, 2012
Manufacturer
HOSPIRA INC.
Product Code
FRN
PMA / PMN Number
K060806
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION IS NOT COMPLETE. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

GENERAL REPORT RECEIVED OF AN UNSPECIFIED NUMBER OF UNDOCUMENTED INCIDENTS OF THE DEVICE DID NOT DELIVER. ON UNSPECIFIED DATES, IT WAS REPORTED THAT GEMSTAR PUMPS AND GEMSTAR BOLUS CORDS WERE CONNECTED TO GEMSTAR DOCKING STATIONS AND WERE BEING USED TO DELIVER UNSPECIFIED MEDICATIONS. NO FURTHER PROGRAMMING PARAMETERS WERE PROVIDED. AFTER UNSPECIFIED LENGTHS OF TIME, IT WAS REPORTED THAT THE BOLUS CORDS WERE NOT WORKING. THE DOCKING STATIONS WERE REPLACED AND THE THERAPIES WERE RESUMED. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICES WERE IN CLINICAL USE. UPON VISUAL EXAMINATIONS IN THE BIOMEDICAL ENGINEERING DEPARTMENT, IT WAS NOTED THAT AN UNSPECIFIED NUMBER OF PINS WERE MISSING FROM THE BOLUS CORD CONNECTOR PORT OF THE DOCKING STATIONS. THOUGH REQUESTED, NO ADDITIONAL INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
23045 GEMSTAR DOCKING STATION 80FRN FRN HOSPIRA INC. NA UNKG1

Patients

Seq Age Sex Outcome Treatment
1 UNK GEMSTAR PUMP: LIST #UNK, SN UNK| GEMSTAR BOLUS CORD: LIST #13026, LOT #UNK