UNKNOWN ZIMMER KNEE
Report
- Report Number
- 1822565-2013-00094
- Event Type
- Injury
- Date Received
- January 15, 2013
- Report Date
- December 17, 2012
- Manufacturer
- ZIMMER, INC.
- Product Code
- HRZ
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
INFORMATION WAS RECEIVED FROM A CONSUMER WHO IS NOT REQUIRED TO COMPLETE FORM 3500A. NO DEVICES OR PHOTOS WERE RECEIVED; THEREFORE, THE CONDITION OF THE COMPONENTS IS UNKNOWN. NEITHER SURGICAL NOTES NOR X-RAYS WERE PROVIDED; THEREFORE, FIT AND ORIENTATION COULD NOT BE EVALUATED. PT FACTORS THAT MAY AFFECT THE PERFORMANCE OF THE COMPONENTS SUCH AS BONE QUALITY, ACTIVITY LEVEL OR TYPE OF ACTIVITY (LOW IMPACT VS. HIGH IMPACT) ARE UNKNOWN. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED WITH THE INFORMATION PROVIDED. REVIEW OF THE DEVICE HISTORY RECORDS WAS NOT POSSIBLE AS THE PRODUCT AND/OR LOT NUMBERS REQUIRED FOR A RETRIEVAL WERE UNAVAILABLE. IT IS NOT SUSPECTED THAT THE PRODUCT FAILED TO MEET SPECIFICATIONS. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY EVIDENCE OF PRODUCT CONTRIBUTION TO THE REPORTED PROBLEM. BASED ON THE AVAILABLE INFORMATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. SHOULD ADDITIONAL SUBSTANTIVE INFORMATION BE RECEIVED, THE COMPLAINT WILL BE REOPENED. ZIMMER, INC. CONSIDERS THE INVESTIGATION CLOSED.
IT IS REPORTED THAT THE PT EXPERIENCED LEFT KNEE SWELLING AND HEARD A CLICKING NOISE COMING FROM HIS KNEE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 22215 | UNKNOWN ZIMMER KNEE | KNEE PROSTHESIS | HRZ | ZIMMER, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR | Other |