EON MINI
Report
- Report Number
- 1627487-2013-12063
- Event Type
- Injury
- Date Received
- January 10, 2013
- Date of Event
- December 13, 2012
- Report Date
- December 14, 2012
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-12192011-003-R
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WV, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
CORRECTION: 1627487-07262012-001-C. THIS IPG SERIAL NUMBER WAS INCLUDED IN FIELD ADVISORIES AND IN A FIELD CORRECTION. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
REFERENCE MFR REPORT # 1627487-2013-12064. IT WAS REPORTED, THE PATIENT EXPERIENCES POCKET HEATING WHEN CHARGING FOR 20 MINUTES. A NEW CHARGER WAS SENT TO THE PATIENT. FOLLOW-UP IS PENDING. ON 08/01/2012 ST. JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PATIENTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PATIENTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WERE EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 15593 | EON MINI | SCS IPG | LGW | ST. JUDE MEDICAL - NEUROMODULATION | 3788 | 3229703 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR | Required Intervention | IMPLANT DATE:| IMPLANT DATE:| SCS ANCHOR: MODEL 1192 (2)| SCS LEAD: MODEL 3186 (2) |