CHARGING SYSTEM
Report
- Report Number
- 1627487-2013-02076
- Event Type
- Injury
- Date Received
- January 10, 2013
- Date of Event
- December 21, 2012
- Report Date
- December 21, 2012
- Manufacturer
- ST. JUDE MEDICAL - NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS CHARGER MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. CORRECTIVE AND PREVENTIVE ACTION (CAPA) WAS PERFORMED. RESULTS: POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 2 OF 2. REF MFR REPORT: 1627487-2013-02075. IT WAS REPORTED THE PT EXPERIENCES IPG POCKET HEATING WHEN STIMULATION IS ON. SHE REPORTED THE HEATING SENSATION WAS WORSE WHILE CHARGING HER IPG. SHE STATED SHE TURNED OFF HER SCS SYSTEM AS THE PAIN WAS TOO UNCOMFORTABLE AND THE PAIN SUBSIDES WHEN SHE TURNED THE SYSTEM OFF. THE PT REPORTED SHE WOULD PREFER TO HAVE HER IPG REPLACED SINCE SHE GETS PAIN RELIEF FROM THE STIMULATION. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME. ON 08/01/2012, ST. JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 14667 | CHARGING SYSTEM | STS CHARGING SYSTEM | LGW | ST. JUDE MEDICAL - NEUROMODULATION | 3721 | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | Required Intervention | SCS LEAD, MODEL 3214| IMPLANT DATE: |