FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2921546 · Received January 18, 2013

Report

Report Number
1416980-2013-01356
Event Type
Malfunction
Date Received
January 18, 2013
Date of Event
January 3, 2013
Report Date
January 3, 2013
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915522
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CI
Reporter Occupation
SERVICE AND TESTING PERSONNEL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE CONDITION WAS CONFIRMED BY SERVICE. "THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT OPENED DURING THE INVESTIGATION OF (B)(4)." THE CAUSE WAS IDENTIFIED TO BE AN EXPIRED BATTERY. "TO CORRECT THIS CONDITION THE MAIN BATTERY WAS REPLACED." IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP WILL BE SENT.

Description of Event or Problem · 1

DURING PRODUCT EVALUATION BY A BAXTER SERVICE TECHNICIAN, A FLO-GARD INFUSION PUMP WAS FOUND TO HAVE AN OUTDATED BATTERY. "THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY". THIS WAS NOT REPORTED BY THE CUSTOMER. "THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THIS EVENT." NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
28440 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1