FDA Adverse Event Summary report: N

CADD-SOLIS

MDR report key: 2921534 · Received November 28, 2012

Report

Report Number
2921534
Date Received
November 28, 2012
Date of Event
October 30, 2012
Report Date
October 31, 2012
Manufacturer
SMITHS MEDICAL ASD, INC.
Product Code
MEA
Report Source
User Facility report
Reporter Location
DC, US
Reporter Occupation
INVALID DATA

Narratives

Description of Event or Problem · 1

WE EXPERIENCED AN EVENT WITH THE VOLUME REMAINING IN A PCA CASSETTE AFTER PUMP WAS DISCONTINUED. NURSE RECORDED 25ML REMAINING ON FLOW SHEET BUT PUMP DISPLAYED 44ML REMAINING, THEREFORE A 19ML DISCREPANCY. THE NURSE CAME UP WITH 25ML'S BY PULLING THE REMAINING DRUG OUT OF CASSETTE USING A 10CC SYRINGE. EVENT LOG FROM PUMP IDENTIFIED 1.9ML PRIMED AND TOTAL OF 54ML DELIVERED TO PATIENT FROM A 100ML BAG. USING PCA EVENT LOG DATA AND DOING MATH MANUALLY, 44.1ML'S SHOULD HAVE BEEN REMAINING IN CASSETTE. THE RESEARCH STATES THAT PUMP WAS WORKING CORRECTLY AND DELIVERY COUNTS WERE ACCURATE. SINCE PCA CASSETTES ARE FILLED IN-HOUSE, THE REPEATER PUMP USED IN THE PHARMACY WAS ALSO EVALUATED. DURING SEVERAL TEST LOADS, IT WAS IDENTIFIED THE FLUID VOLUME FROM REPEATER PUMP TO FILL CASSETTES WAS INCONSISTENT AND WOULD UNDER-FILL AT RANDOM INTERVALS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CADD-SOLIS PUMP, INFUSION, PCA MEA SMITHS MEDICAL ASD, INC. SOLIS 2110 *
2 REPEATER PUMP REPEATER PUMP MEA SMITHS MEDICAL ASD, INC. 099 *
3 * PUMP, INFUSION, PCA, CASSETTE MEA SMITHS MEDICAL ASD, INC. UNK DRW10/24/12 19:24

Patients

Seq Age Sex Outcome Treatment
1 *