FDA Adverse Event Death Summary report: N

EASY CAP

MDR report key: 29214 · Received December 12, 1995

Report

Report Number
29214
Event Type
Death
Date Received
December 12, 1995
Date of Event
November 5, 1995
Report Date
November 13, 1995
Manufacturer
NELLCOR-PURITAN BENNET, INC.
Product Code
CCK
Adverse Event
Yes
Product Problem
Yes
Report Source
User Facility report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

A 82-YR-OLD MALE ADMITTED TO ER ON 11/5/95 COMPLAINING OF DIZZINESS AND EPIGASTRIC PAIN. SHORTLY AFTER ARRIVAL HE SUFFERED A SEIZURE, DEVELOPED RESPIRATORY DISTRESS REQUIRING INTUBATION. CO2 DETECTOR USED TO CHECK TUBE PLACEMENT. FIRST DEVICE DID NOT WORK. SECOND ONE REVEALED CORRECT PLACEMENT OF ET TUBE. WHILE BAGGING WITH AMBU BAG, VENTILATION BECAME DIFFICULT. PT SUSTAINED BILATERAL PNEUMOTHORACES, REQUIRING CHEST TUBES. PT VENTILATING WELL ON VENTILATOR, HOWEVER DIFFICULTY CONTINUED WITH AMBU BAG. IT WAS ULTIMATELY DISCOVERED THAT WHITE PLASTIC CAP HAD BECOME LODGED IN THE EXHALATION VALVE OF THE AMBU BAG. IT IS BELIEVED THAT THE CAP CAME FROM ONE OF THE DEVICES THE PT EXPIRED TWO DAYS LATER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 EASY CAP END TIDAL CO2 DETECTOR CCK NELLCOR-PURITAN BENNET, INC.

Patients

Seq Age Sex Outcome Treatment
1 82 YR Death